Error Codes
5-3
September 2004
E6: Cycle Skip, Accel.
Problem:
The acceleration exceeds the limits programmed in the
SETUP-SIGNL
submenu of the
User Program
.
Cause:
This condition is usually caused by poor flow conditions
or improper transducer alignment.
Action:
Refer to Chapter 6,
, to correct any flowcell
and/or transducer problems.
E7: Analog Out Error
Problem:
The current in the analog output circuit exceeds the limits
for the analog output port.
Cause:
The output load exceeds the specified limits for the
analog output port.
Action:
Make sure the output load is <600 ohms for the analog
outputs. Check wiring.
E8: Unassigned
E9: Totalizer Overflow
Problem:
The totalizers are unable to keep up with the total
accumulated flow signals.
Cause:
The programmed units/pulse value is too small.
Action:
Select a larger number of units/pulse value or decrease
PULSE ON TIME
.
E10: Unassigned
E11: Unassigned
E12: Unassigned
E13: Volumetric Overflow
Problem:
This error code message indicates that the present
measurement exceeds the range of the meter.
Cause:
A mathematical overflow has occurred in either the
volumetric or mass flow calculations.
Action:
Select larger measurement units or a shorter time interval
for the current measurement parameter. For example,
choose
MGAL/M
instead of
GAL/M
in the
CHx-SYSTM
menu (see page 2-3).
Содержание AquaTrans AT868
Страница 8: ...Chapter 1 ...
Страница 24: ...Chapter 2 ...
Страница 60: ...Chapter 3 ...
Страница 66: ...Chapter 4 ...
Страница 72: ...Chapter 5 ...
Страница 77: ...Chapter 6 ...
Страница 88: ...Chapter 7 ...
Страница 98: ...Chapter 8 ...
Страница 106: ...Appendix A ...
Страница 112: ...Appendix B ...
Страница 113: ...CE Mark Compliance Introduction B 1 EMC Compliance B 1 LVD Compliance B 2 ...
Страница 116: ...Appendix C ...
Страница 117: ...Data Records Site Data C 1 ...
Страница 122: ...Appendix D ...
Страница 123: ...Measuring P and L Dimensions Measuring P and L D 1 ...
Страница 126: ...Appendix E ...
Страница 127: ...Service Record Introduction E 1 Data Entry E 1 Diagnostic Parameters E 3 ...
Страница 129: ...September 2004 E 2 Service Record Table E 1 Service Record cont Date Description of Service Performed By ...
Страница 131: ...Appendix F ...