Diagnostics
6-3
September 2004
Procedure Options
After completing the above steps, the meter returns to the
Global I/O
window. Do one of the following:
•
To perform another programming function, see Appendix A,
, to navigate to the desired menu.
•
To leave the
User Program
, press
[ESC]
four times.
After leaving the
User Program
, the AT868 will reset itself and will
begin to display the parameters specified in this section. If more than
one parameter was set up, each of the parameters will be displayed in
sequence, with a pause of several seconds between display changes.
NFup
Displays the normalization factor for the
upstream transducer.
0.85-1.0
<0.85
NFdn
Displays the normalization factor for the
downstream transducer.
0.85-1.0
<0.85
Cxdcr
Displays the speed of sound in the
transducer (clamp-on only).
N.A.
N.A.
Table 6-1: Available Diagnostic Parameters (Continued)
Option Bar
Description
Good
Bad
Содержание AquaTrans AT868
Страница 8: ...Chapter 1 ...
Страница 24: ...Chapter 2 ...
Страница 60: ...Chapter 3 ...
Страница 66: ...Chapter 4 ...
Страница 72: ...Chapter 5 ...
Страница 77: ...Chapter 6 ...
Страница 88: ...Chapter 7 ...
Страница 98: ...Chapter 8 ...
Страница 106: ...Appendix A ...
Страница 112: ...Appendix B ...
Страница 113: ...CE Mark Compliance Introduction B 1 EMC Compliance B 1 LVD Compliance B 2 ...
Страница 116: ...Appendix C ...
Страница 117: ...Data Records Site Data C 1 ...
Страница 122: ...Appendix D ...
Страница 123: ...Measuring P and L Dimensions Measuring P and L D 1 ...
Страница 126: ...Appendix E ...
Страница 127: ...Service Record Introduction E 1 Data Entry E 1 Diagnostic Parameters E 3 ...
Страница 129: ...September 2004 E 2 Service Record Table E 1 Service Record cont Date Description of Service Performed By ...
Страница 131: ...Appendix F ...