Remedies
Possible causes
Problems
•
Check all power connections
•
Check buffer level in tanks
•
Ensure that the gel cassette
is in place
•
Ensure that there is no tape
left on the gel cassette
Incomplete circuit
No green light, low
current (<20 mA)
at start.
•
Check that the tanks are not
overfilled
•
Check if the unit is damaged
and needs to be replaced
Leak between tanks, with or
without gel cassette
Green light, but
only approx. 35
mA at start.
•
Check that tanks are not
overfilled
•
Check if the unit is damaged
and needs to be replaced
Leak between tanks with gel
cassette
Green light, but
approx. 65 mA at
start
Amersham ECL Gel and Gel Box User Manual 28-9943-33 AC
61
7 Troubleshooting