Preparing the transfer tank
Action
Step
Fill the transfer unit with transfer buffer.
1
For information on how to use the transfer unit, refer to the
instructions provided by the manufacturer of the transfer unit.
Place the open transfer cassette with the anode side (+) down in
a tray with transferbuffer and ensure that it is immersed in at
least 3 cm of the transfer buffer.
2
Build the stack with the membrane closest to the anode side.
3
Below is an illustration of the transfer sandwich.
CATHODE
ANODE
Transfer dir
ection
Two buffer soaked
blotting papers
Foam Sponge
Foam Sponge
Two buffer soaked
blotting papers
Polyacrylamide Gel
Membrane
Layer 1: Place a 3 mm thick foam sponge in the transfer cassette
and press gently to remove any air bubbles.
4
Wet the blotting paper in transfer buffer.
5
54
Amersham ECL Gel and Gel Box User Manual 28-9943-33 AC
5 Protocols
5.4 Transfer protocol