Power supply
Amersham ECL Gel Box is powered by an external power supply (EPS 301).
See
for more information. If other power supply
with a Ø 4 mm, deep socket (>6 mm internal distance between surface and
socket) is used, it is recommended to use the adapter delivered with
Amersham ECL Gel Box. See
Connecting other power supplies, on page 35
for more information.
Safety lid
Amersham ECL Gel Box is equipped with a safety lid, which shall be placed
on top of the base unit before a power supply is connected.
The safety lid has two magnets which will activate current only when their
magnetic fields are in contact with the electrodes inside the base unit of
Amersham ECL Gel Box. This means that Amersham ECL Gel Box is only
powered when the safety lid is in the correct position. Removing the safety
lid will switch off the high voltage to the electrodes.
WARNING
Do not operate Amersham ECL Gel Box without the
safety lid.
Amersham ECL Gel and Gel Box User Manual 28-9943-33 AC
29
3 System description
3.1 Amersham ECL Gel Box