TROUBLESHOOTING: Service Menus
Revision B
Dash 3000/4000 Patient Monitor
5-5
2000966-105
Access the Service Mode
Access the SERVICE MODE menu starting from the MAIN menu.
1.
Select MORE MENUS -> MONITOR SETUP -> SERVICE MODE ->
2.
Enter password using the Trim Knob control to select the day and
month from monitor screen with leading zeros. (e.g. July 4 = 0407).
About Service Mode
Menu Option Items
The Service Mode menu is used for initial setup and configuration as well
as for troubleshooting. ALWAYS exercise caution when using any of
these password-protected functions.
The service technician can use the Service Mode menu to:
◆
relay software information to design engineers;
◆
calibrate and troubleshoot NBP functions of the monitor;
◆
set admit menu options, operating mode of the monitor, and
monitor default password;
◆
configure the monitor unit name, bed number and Internet
address for use on the network; and
◆
enter or change the time and date on the monitor.
Do not use any of these options unless specifically instructed to do so.
WARNING
Some of the service mode menu option items are to be
used only by qualified service technicians and others are
for general use. Because of this, unnecessary tampering
with service mode menu option items for experimentation
purposes is not recommended by GE and may cause a
malfunction of the monitor.
Service Mode Menu
Option Items
Following is a list of options in the main code service menu;
REVIEW ERRORS—This menu selection is for advanced troubleshooting
by GE engineers. Error log data can be transferred over the network to a
central station and then loaded onto a diskette for review. (Review Errors
is discusses in greater detail later in this chapter.)
CALIBRATE—For checkout or calibration of the noninvasive blood
pressure, ECG analog output, BP analog output, CO
2
service, and SAM
service menu functions of the monitor.
BATTERY SERVICE—This is a complete collection of battery data for
troubleshooting the batteries.
030B
Содержание Dash 3000
Страница 11: ...Revision B Dash 3000 4000 Patient Monitor 1 1 2000966 105 1 INTRODUCTION ...
Страница 12: ...INTRODUCTION 1 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 19: ...Revision B Dash 3000 4000 Patient Monitor 2 1 2000966 105 2 EQUIPMENT OVERVIEW ...
Страница 20: ...EQUIPMENT OVERVIEW 2 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 43: ...Revision B Dash 3000 4000 Patient Monitor 3 1 2000966 105 3 INSTALLATION ...
Страница 44: ...INSTALLATION 3 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 51: ...Revision B Dash 3000 4000 Patient Monitor 4 1 2000966 105 4 MAINTENANCE ...
Страница 52: ...MAINTENANCE 4 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 93: ...Revision B Dash 3000 4000 Patient Monitor 5 1 2000966 105 5 TROUBLESHOOTING ...
Страница 94: ...TROUBLESHOOTING 5 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 122: ...TROUBLESHOOTING Language Specific Information 5 30 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 123: ...Revision B Dash 3000 4000 Patient Monitor 6 1 2000966 105 6 CONFIGURATION ...
Страница 124: ...CONFIGURATION 6 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 144: ...CONFIGURATION 6 22 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 145: ...Revision B Dash 3000 4000 Patient Monitor 7 1 2000966 105 7 CALIBRATION ...
Страница 146: ...CALIBRATION 7 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 154: ...FIELD REPLACEABLE UNITS AND UPGRADES 8 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 189: ...Revision B Dash 3000 4000 Patient Monitor 9 1 2000966 105 9 ASSEMBLY DRAWINGS ...
Страница 190: ...ASSEMBLY DRAWINGS 9 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 201: ...ASSEMBLY DRAWINGS Theory Of Operation Revision B Dash 3000 4000 Patient Monitor 9 13 2000966 105 Block Diagram ...
Страница 230: ...ASSEMBLY DRAWINGS Port Connections 9 42 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 231: ......
Страница 232: ...GE Medical Systems Information Technologies g gemedicalsystem com ...