EQUIPMENT OVERVIEW: Technical Specifications
Revision B
Dash 3000/4000 Patient Monitor
2-19
2000966-105
Analog Output
Defibrillator
Synchronization
Pulse
ECG:
Gain:
DC offset:
Noise:
Frequency response:
Time delay:
1 V/mV ±10%
±100 mV (max)
<5 mVp-p (0-300 Hz)
In the “ECG” section, see the “Output specifications:”
on page 2-12
40 ms monitoring filter, 35 ms diagnostic filter
Blood pressure:
Gain:
DC offset:
Noise:
Frequency response:
Time delay:
10 mV/mmHg ±2%
±20 mV (max)
<5 mVp-p (0-300 Hz)
dc to 50 Hz +2/–0 Hz
40 Hz filter, 37 ms
Marker out:
Time delay:
Amplitude (selectable in Service
menu):
+5 V selection:
+12 V selection:
Pulse width:
Output impedance:
Current limit:
35 ms (maximum), R-wave peak to leading edge of
pulse.
3.5 V (min) at 1 mA sourcing; 0.5 V (max) at 5 mA
sinking.
11.0 V (min) at 1 mA sourcing; 0.75 V (max) at 5 mA
sinking.
10 ms ±10% or 100 ms ±10% (selectable in Service
menu).
50
Ω
nominal
15 mA nominal, both sourcing and sinking.
Marker in:
Input threshold:
Input hysteresis:
Maximum input voltage:
Input impedance:
Pulse width:
V
IH
= ±2.5 V (min); V
IL
= ±1.5 V (max)
650 mV typical
±30 V (with respect to ground on pin 3)
10 k
Ω
(min) for -25 V < VIN < 25 V
1.0 ms (min), VIN > 2.5 V
Содержание Dash 3000
Страница 11: ...Revision B Dash 3000 4000 Patient Monitor 1 1 2000966 105 1 INTRODUCTION ...
Страница 12: ...INTRODUCTION 1 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 19: ...Revision B Dash 3000 4000 Patient Monitor 2 1 2000966 105 2 EQUIPMENT OVERVIEW ...
Страница 20: ...EQUIPMENT OVERVIEW 2 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 43: ...Revision B Dash 3000 4000 Patient Monitor 3 1 2000966 105 3 INSTALLATION ...
Страница 44: ...INSTALLATION 3 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 51: ...Revision B Dash 3000 4000 Patient Monitor 4 1 2000966 105 4 MAINTENANCE ...
Страница 52: ...MAINTENANCE 4 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 93: ...Revision B Dash 3000 4000 Patient Monitor 5 1 2000966 105 5 TROUBLESHOOTING ...
Страница 94: ...TROUBLESHOOTING 5 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 122: ...TROUBLESHOOTING Language Specific Information 5 30 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 123: ...Revision B Dash 3000 4000 Patient Monitor 6 1 2000966 105 6 CONFIGURATION ...
Страница 124: ...CONFIGURATION 6 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 144: ...CONFIGURATION 6 22 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 145: ...Revision B Dash 3000 4000 Patient Monitor 7 1 2000966 105 7 CALIBRATION ...
Страница 146: ...CALIBRATION 7 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 154: ...FIELD REPLACEABLE UNITS AND UPGRADES 8 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 189: ...Revision B Dash 3000 4000 Patient Monitor 9 1 2000966 105 9 ASSEMBLY DRAWINGS ...
Страница 190: ...ASSEMBLY DRAWINGS 9 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 201: ...ASSEMBLY DRAWINGS Theory Of Operation Revision B Dash 3000 4000 Patient Monitor 9 13 2000966 105 Block Diagram ...
Страница 230: ...ASSEMBLY DRAWINGS Port Connections 9 42 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 231: ......
Страница 232: ...GE Medical Systems Information Technologies g gemedicalsystem com ...