CALIBRATION: NBP, Analog Output ECG, Analog Output BP, and End-tidal CO2 Software Calibration
Revision B
Dash 3000/4000 Patient Monitor
7-5
2000966-105
The table below lists items for connecting the NBP tube between the
manometer and NBP cuff:
WARNING
When the NBP cuff is used in this procedure, it must be
tightly wrapped around a rigid cylinder or pipe. DO NOT
put the NBP cuff around a human arm during the
calibration procedures due to the potential for injury.
Calibration Procedure
Using the Trim Knob control, access the SERVICE MODE menu starting
from the MAIN menu.
1.
Select MORE MENUS-> MONITOR SETUP-> SERVICE MODE.
2.
Enter password using the Trim Knob control to select the day and
month from monitor screen with leading zeros. (e.g. July 4 = 0407).
3.
Select CALIBRATE-> CALIBRATE NBP-> CAL ZERO OFF->
START.
4.
The text on the menu item changes from CAL ZERO OFF to CAL
ZERO IN PROGRESS.
When the process is complete, the menu item shows that it is OFF
again.
4
To proceed with the Gain Calibration Test, setup the monitor and test
equipment following the guidelines illustrated in the NBP Checkout
Procedure, Chapter 4, Maintenance.
Description
Part Number
Qty
NBP cuff coupling
400787-001
1
NBP hose coupling
46100-002
1
NBP tee
4745-101
1
NBP tubing
401582-001
2
Содержание Dash 3000
Страница 11: ...Revision B Dash 3000 4000 Patient Monitor 1 1 2000966 105 1 INTRODUCTION ...
Страница 12: ...INTRODUCTION 1 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 19: ...Revision B Dash 3000 4000 Patient Monitor 2 1 2000966 105 2 EQUIPMENT OVERVIEW ...
Страница 20: ...EQUIPMENT OVERVIEW 2 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 43: ...Revision B Dash 3000 4000 Patient Monitor 3 1 2000966 105 3 INSTALLATION ...
Страница 44: ...INSTALLATION 3 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 51: ...Revision B Dash 3000 4000 Patient Monitor 4 1 2000966 105 4 MAINTENANCE ...
Страница 52: ...MAINTENANCE 4 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 93: ...Revision B Dash 3000 4000 Patient Monitor 5 1 2000966 105 5 TROUBLESHOOTING ...
Страница 94: ...TROUBLESHOOTING 5 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 122: ...TROUBLESHOOTING Language Specific Information 5 30 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 123: ...Revision B Dash 3000 4000 Patient Monitor 6 1 2000966 105 6 CONFIGURATION ...
Страница 124: ...CONFIGURATION 6 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 144: ...CONFIGURATION 6 22 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 145: ...Revision B Dash 3000 4000 Patient Monitor 7 1 2000966 105 7 CALIBRATION ...
Страница 146: ...CALIBRATION 7 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 154: ...FIELD REPLACEABLE UNITS AND UPGRADES 8 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 189: ...Revision B Dash 3000 4000 Patient Monitor 9 1 2000966 105 9 ASSEMBLY DRAWINGS ...
Страница 190: ...ASSEMBLY DRAWINGS 9 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 201: ...ASSEMBLY DRAWINGS Theory Of Operation Revision B Dash 3000 4000 Patient Monitor 9 13 2000966 105 Block Diagram ...
Страница 230: ...ASSEMBLY DRAWINGS Port Connections 9 42 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 231: ......
Страница 232: ...GE Medical Systems Information Technologies g gemedicalsystem com ...