TROUBLESHOOTING: Data Acquisition Tests
5-16
Dash 3000/4000 Patient Monitor
Revision B
2000966-105
Pace Detect
Functions
1.
With the monitor displaying leads I, II, III, and V, set the patient
simulator to output a VP1 (ventricular pacemaker simulation #1)
waveform.
2.
Enable the pacemaker detection function of the monitor:
•
select ECG from the display main menu,
•
select DETECT PACE and set to PACE 2.
3.
Verify the heart rate remains at approximately 80 bpm.
4.
Select the VP2 output (ventricular pacemaker simulation #2) on the
patient simulator.
5.
Select and press RELEARN.
6.
The heart rate number may disappear from the display for a few
seconds and return to the screen shortly thereafter. Verify the heart
rate is at approximately 80 bpm. Verify the pacemaker spikes display
at the same amplitude.
7.
Disable the pacemaker detection function of the monitor. Verify the
displayed pacemaker spikes have a different amplitude than in the
previous step.
8.
Select the AVS output (A/V sequential pacemaker simulation) on the
patient simulator.
9.
Select and press RELEARN.
10. Verify the displayed pacemaker spikes are at different amplitudes.
11. Select and press RELEARN.
12. Enable the pacemaker detection function of the monitor once again.
13. Set ECG GAIN to X2.
14. Verify a stable heart rate display of approximately 80 bpm. Verify the
pacemaker spikes are again at the same amplitude.
15. Disable the pacemaker detection function of the monitor.
Pace Detect Functions Do
Not Work Properly
If the pacemaker detection test results are not correct, as described
above:
•
Verify the patient simulator functions correctly by testing it on a
working monitor,
•
The acquisition PCB is suspect. Swap a working acquisition PCB into
the monitor and perform this test to verify correct operation.
Содержание Dash 3000
Страница 11: ...Revision B Dash 3000 4000 Patient Monitor 1 1 2000966 105 1 INTRODUCTION ...
Страница 12: ...INTRODUCTION 1 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 19: ...Revision B Dash 3000 4000 Patient Monitor 2 1 2000966 105 2 EQUIPMENT OVERVIEW ...
Страница 20: ...EQUIPMENT OVERVIEW 2 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 43: ...Revision B Dash 3000 4000 Patient Monitor 3 1 2000966 105 3 INSTALLATION ...
Страница 44: ...INSTALLATION 3 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 51: ...Revision B Dash 3000 4000 Patient Monitor 4 1 2000966 105 4 MAINTENANCE ...
Страница 52: ...MAINTENANCE 4 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 93: ...Revision B Dash 3000 4000 Patient Monitor 5 1 2000966 105 5 TROUBLESHOOTING ...
Страница 94: ...TROUBLESHOOTING 5 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 122: ...TROUBLESHOOTING Language Specific Information 5 30 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 123: ...Revision B Dash 3000 4000 Patient Monitor 6 1 2000966 105 6 CONFIGURATION ...
Страница 124: ...CONFIGURATION 6 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 144: ...CONFIGURATION 6 22 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 145: ...Revision B Dash 3000 4000 Patient Monitor 7 1 2000966 105 7 CALIBRATION ...
Страница 146: ...CALIBRATION 7 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 154: ...FIELD REPLACEABLE UNITS AND UPGRADES 8 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 189: ...Revision B Dash 3000 4000 Patient Monitor 9 1 2000966 105 9 ASSEMBLY DRAWINGS ...
Страница 190: ...ASSEMBLY DRAWINGS 9 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 201: ...ASSEMBLY DRAWINGS Theory Of Operation Revision B Dash 3000 4000 Patient Monitor 9 13 2000966 105 Block Diagram ...
Страница 230: ...ASSEMBLY DRAWINGS Port Connections 9 42 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 231: ......
Страница 232: ...GE Medical Systems Information Technologies g gemedicalsystem com ...