TROUBLESHOOTING: Data Acquisition Tests
Revision B
Dash 3000/4000 Patient Monitor
5-21
2000966-105
Noninvasive Blood
Pressure Functions
Perform the noninvasive blood pressure (NBP) Checkout Procedure
found in the Maintenance chapter. This procedure determines whether or
not the NBP functions of the monitor are working as designed or whether
the monitor requires NBP calibration.
If, after performing the prescribed checkout procedure, it is determined
that there are potential problems that NBP calibration does not cure, try
the following:
1.
If calibration is unsuccessful and cannot be properly performed,
there could be leaks in the pneumatic circuit plumbing. The following
steps assist you in determining this:
•
The NBP cuff and tubing is the easiest area to inspect for leaks
and is also the most likely area for failure in this regard. Closely
inspect these items for cracks or leaks. Test the NBP cuff and
tubing on a working monitor to determine the source of the
problem.
•
If the NBP cuff and tubing are determined to be good after
testing them on a working monitor, the leaks are probably
internal to the monitor. Disassemble the monitor and check
inspect all internal tubing and connections in the pneumatic
circuit plumbing.
2.
If no leaks are found after performing the previous step, the NBP
pump assembly is suspect. Swap the NBP pump assembly with one
from a working monitor and/or replace as necessary.
NBP Alarms Occur
Continuously
If the monitor is not configured properly, a variety of NBP problems may
occur. To determine monitor configuration, rotate then push the Trim
Knob control to:
•
Scroll to and select CUSTOMIZE MONITOR from the main menu of
the monitor,
•
Scroll to and select PATIENT-MONITOR TYPE from the customize
monitor menu of the monitor.
Verify the configured monitor type matches the environment in which
the monitor is being used. If it is set to a neonatal ICU when the monitor
is used for the adult ICU application or vice versa, problems listed to the
left may occur.
Содержание Dash 3000
Страница 11: ...Revision B Dash 3000 4000 Patient Monitor 1 1 2000966 105 1 INTRODUCTION ...
Страница 12: ...INTRODUCTION 1 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 19: ...Revision B Dash 3000 4000 Patient Monitor 2 1 2000966 105 2 EQUIPMENT OVERVIEW ...
Страница 20: ...EQUIPMENT OVERVIEW 2 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 43: ...Revision B Dash 3000 4000 Patient Monitor 3 1 2000966 105 3 INSTALLATION ...
Страница 44: ...INSTALLATION 3 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 51: ...Revision B Dash 3000 4000 Patient Monitor 4 1 2000966 105 4 MAINTENANCE ...
Страница 52: ...MAINTENANCE 4 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 93: ...Revision B Dash 3000 4000 Patient Monitor 5 1 2000966 105 5 TROUBLESHOOTING ...
Страница 94: ...TROUBLESHOOTING 5 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 122: ...TROUBLESHOOTING Language Specific Information 5 30 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 123: ...Revision B Dash 3000 4000 Patient Monitor 6 1 2000966 105 6 CONFIGURATION ...
Страница 124: ...CONFIGURATION 6 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 144: ...CONFIGURATION 6 22 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 145: ...Revision B Dash 3000 4000 Patient Monitor 7 1 2000966 105 7 CALIBRATION ...
Страница 146: ...CALIBRATION 7 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 154: ...FIELD REPLACEABLE UNITS AND UPGRADES 8 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 189: ...Revision B Dash 3000 4000 Patient Monitor 9 1 2000966 105 9 ASSEMBLY DRAWINGS ...
Страница 190: ...ASSEMBLY DRAWINGS 9 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 201: ...ASSEMBLY DRAWINGS Theory Of Operation Revision B Dash 3000 4000 Patient Monitor 9 13 2000966 105 Block Diagram ...
Страница 230: ...ASSEMBLY DRAWINGS Port Connections 9 42 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 231: ......
Страница 232: ...GE Medical Systems Information Technologies g gemedicalsystem com ...