MAINTENANCE: Checkout Procedures
Revision B
Dash 3000/4000 Patient Monitor
4-29
2000966-105
8.
Observe a distortion-free AR1 BP waveform and a reading of
approximately 120/80 (93) in the AR1 parameter window on the
monitor display.
9.
Disconnect the BP simulator cable from the BP1 connector of the
monitor. Continue to the next step for the BP2 test.
10. Again, set up the patient simulator as follows:
•
BP polarity – POS,
•
BP output – 0 mmHg.
BP2 Connector (PA2)
Tests
1.
Connect the BP simulator cable to the BP2 (right-most BP)
connector of the monitor.
2.
Verify a PA2 parameter window, waveform label and corresponding
graticules appear on the monitor display, along with a PA2 BP
waveform requiring zero reference.
3.
Press the ZERO ALL key on the front panel of the monitor to zero
reference the PA2 BP waveform.
4.
Change the patient simulator BP output to 200 mmHg (240 if using
Marq II).
5.
Observe a reading of 200/200 (200) ± 4 mmHg (or 240 ± 4 mmHg if
using Marq II) in the PA2 parameter window on the monitor display.
6.
Change the patient simulator BP output to WAVE (simulated BP
waveform).
7.
Set the PA2 BP waveform gain on the monitor to auto.
8.
Observe a distortion-free PA2 BP waveform and a reading of
approximately 120/80 (93) in the PA2 parameter window on the
monitor display.
9.
Remove the BP simulator cable from the BP2 connector of the
monitor. This completes the BP tests.
Содержание Dash 3000
Страница 11: ...Revision B Dash 3000 4000 Patient Monitor 1 1 2000966 105 1 INTRODUCTION ...
Страница 12: ...INTRODUCTION 1 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 19: ...Revision B Dash 3000 4000 Patient Monitor 2 1 2000966 105 2 EQUIPMENT OVERVIEW ...
Страница 20: ...EQUIPMENT OVERVIEW 2 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 43: ...Revision B Dash 3000 4000 Patient Monitor 3 1 2000966 105 3 INSTALLATION ...
Страница 44: ...INSTALLATION 3 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 51: ...Revision B Dash 3000 4000 Patient Monitor 4 1 2000966 105 4 MAINTENANCE ...
Страница 52: ...MAINTENANCE 4 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 93: ...Revision B Dash 3000 4000 Patient Monitor 5 1 2000966 105 5 TROUBLESHOOTING ...
Страница 94: ...TROUBLESHOOTING 5 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 122: ...TROUBLESHOOTING Language Specific Information 5 30 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 123: ...Revision B Dash 3000 4000 Patient Monitor 6 1 2000966 105 6 CONFIGURATION ...
Страница 124: ...CONFIGURATION 6 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 144: ...CONFIGURATION 6 22 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 145: ...Revision B Dash 3000 4000 Patient Monitor 7 1 2000966 105 7 CALIBRATION ...
Страница 146: ...CALIBRATION 7 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 154: ...FIELD REPLACEABLE UNITS AND UPGRADES 8 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 189: ...Revision B Dash 3000 4000 Patient Monitor 9 1 2000966 105 9 ASSEMBLY DRAWINGS ...
Страница 190: ...ASSEMBLY DRAWINGS 9 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 201: ...ASSEMBLY DRAWINGS Theory Of Operation Revision B Dash 3000 4000 Patient Monitor 9 13 2000966 105 Block Diagram ...
Страница 230: ...ASSEMBLY DRAWINGS Port Connections 9 42 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Страница 231: ......
Страница 232: ...GE Medical Systems Information Technologies g gemedicalsystem com ...