Module Frames and Modules
9 - 30
2062973-004
Monitor configuration
•
Configure the Flow waveform field to the monitor screen with adequate priority.
•
Select the
Setup
tab in the
Spirometry and Gas Exchange
menu and configure:
Scaling:
Auto
Sensor Type:
Adult
Show Volume:
TV
Ventilator configuration
•
Configure the ventilator to use air as fresh gas.
•
Set the Tidal Volume (TV) to 500 ml/min when doing calibration check and calibration with
adult sensor and 100 ml/min with pediatric sensor.
•
Set the RR =15, I/E =1/2 and PEEP 0cmH2O.
4.3.2 Calibration check
1.
Perform the calibration check according to the steps 1 through 12a in the "Instructions for
Use" -letter of the spirometry tester.
NOTE: Let the gas module to warm up at least for 10 minutes before performing the calibration
check or flow calibration.
2.
The measured flow values are shown in real-time in the
TV Insp
and
TV Exp
fields in the
Flow parameter window. Compare these measured values to the TV value reading
(highest water level) in the spirometry tester.
Acceptance criteria:
•
If the
TV Insp
and
TV Exp
values differ less than ± 6 % of the value read from the
spirometry tester, flow calibration is not needed.
•
If the
TV Insp
and
TV Exp
values differ more than ± 6 % of the value read from the
spirometry tester, perform flow calibration according to section 4.3.3 Flow calibration.
4.3.3 Flow calibration
1.
Select
Monitor Setup
>
Service Calibrations
2.
Enter the User Name and the Password and press
Enter
to get into the Calibrations menu.
3.
Select
Gases
4.
Select
Spirometry Calibration
5.
Ensure that the
Sensor Type
is correct and that
Spirometry Zeroing
is Enabled.
6.
Wait until the PVX sensor performs an automatic zeroing. It will show a message "zeroing"
in the Flow parameter window when zeroing takes place.
7.
Adjust the
Exp Flow Gain
and/or
Insp Flow Gain
separately by using the related up-down
spinner controls to calibrate the measured
TV Exp ml
and
TV Insp ml
values:
−
To increase the
TV Exp ml
flow value, increase the
Exp Flow Gain
.
−
To decrease the
TV Exp ml
flow value, lower the
Exp Flow Gain
.
−
To increase the
TV Insp ml
flow value, increase the
TV Flow Gain
.
−
To decrease the
TV Insp ml
flow value, lower the
TV Flow Gain
.
8.
Press
Confirm
to check the effect of the gain adjustment to the flow readings.
9.
Repeat steps 6 and 7 until the flow values are within the specification.
Содержание E-PSMP
Страница 2: ......
Страница 4: ...Module Frames and Modules 2062973 004 For your notes...
Страница 6: ...Module Frames and Modules 1 ii 2062973 004...
Страница 10: ...Module Frames and Modules 1 4 2062973 004 For your notes...
Страница 11: ...2 Hemodynamic Modules E PRESTN E RESTN E PRETN Rev 00 E PSMP E PSM Rev 01 Technical Manual...
Страница 12: ......
Страница 92: ...Module Frames and Modules 2 78 2062973 004 For your notes...
Страница 94: ...Module Frames and Modules 2 80 2062973 004 For your notes...
Страница 95: ...3 Continuous Cardiac Output Module E PiCCO Technical Manual...
Страница 96: ......
Страница 124: ...Module Frames and Modules 3 26 2062973 004 For your notes...
Страница 125: ...4 Cardiac Output and SvO2 Module E COPSv Rev 01 Cardiac Output Module E COP Rev 01 Technical Manual...
Страница 126: ......
Страница 138: ...Module Frames and Modules 4 10 2062973 004 2 Configuration There is no configuration for the E COP or E COPSv modules...
Страница 159: ...Cardiac Output Modules E COPSv E COP 4 31 2062973 004...
Страница 160: ...Module Frames and Modules 4 32 2062973 004 For your notes...
Страница 162: ...Module Frames and Modules 4 34 2062973 004 For your notes...
Страница 163: ...5 Dual Pressure Module E PP Pressure Temp Module E PT Pressure Module E P Technical Manual...
Страница 164: ......
Страница 198: ...Module Frames and Modules 5 32 2062973 004 For your notes...
Страница 199: ...6 Masimo Module E MASIMO Technical Manual...
Страница 200: ......
Страница 202: ...Module Frames and Modules 6 ii 2062973 004...
Страница 207: ...Masimo Module E MASIMO 6 5 2062973 004 2 Configuration There is no configuration for the E MASIMO module...
Страница 218: ...Module Frames and Modules 6 16 2062973 004 For your notes...
Страница 220: ...Module Frames and Modules 6 18 2062973 004 For your notes...
Страница 221: ...7 Nellcor Compatible Module E NSAT E NSATX Technical Manual...
Страница 222: ......
Страница 224: ...Module Frames and Modules 7 ii 2062973 004...
Страница 236: ...Module Frames and Modules 7 12 2062973 004...
Страница 244: ...Module Frames and Modules 7 20 2062973 004 For your notes...
Страница 246: ...Module Frames and Modules 7 22 2062973 004 For your notes...
Страница 247: ...8 Respiratory Modules E sCAiOVX E sCAiOV E sCAiO E sCOVX E sCOV E sCO Technical Manual...
Страница 248: ......
Страница 284: ...Module Frames and Modules 8 34 2062973 004...
Страница 316: ...Module Frames and Modules 8 66 2062973 004 For your notes...
Страница 319: ...9 Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV E CO Technical Manual...
Страница 320: ......
Страница 336: ...Module Frames and Modules 9 14 2062973 004 For your notes...
Страница 353: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 31 2062973 004...
Страница 379: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 57 2062973 004...
Страница 380: ...Module Frames and Modules 9 58 2062973 004 For your notes...
Страница 381: ...10 Single width Airway Module E miniC Technical Manual...
Страница 382: ......
Страница 391: ...Single width Airway Module E miniC 10 9 2062973 004 2 Configuration There is no configuration for the E miniC module...
Страница 397: ...Single width Airway Module E miniC 10 15 2062973 004 Complete on the Maintenance check form on page 10 36...
Страница 410: ...Module Frames and Modules 10 28 2062973 004...
Страница 419: ...11 Entropy Module E ENTROPY Rev 01 Technical Manual...
Страница 420: ......
Страница 422: ...Module Frames and Modules 11 ii 2062973 004 7 2 Entropy Module E ENTROPY 01 11 17 Maintenance check form 11 19...
Страница 427: ...Entropy Module E ENTROPY Rev 01 11 5 2062973 004 2 Configuration There is no configuration for the E ENTROPY module...
Страница 442: ...Module Frames and Modules 11 20 2062973 004 For your notes...
Страница 443: ...12 EEG Module E EEG and EEG Headbox N EEG Technical Manual...
Страница 444: ......
Страница 446: ...Module Frames and Modules 12 ii 2062973 004 7 3 EEG Headbox N EEG 12 20 Maintenance check form 12 23...
Страница 452: ...Module Frames and Modules 12 6 2062973 004 2 Configuration There is no configuration for the E EEG module...
Страница 468: ...Module Frames and Modules 12 22 2062973 004 For your notes...
Страница 470: ...Module Frames and Modules 12 24 2062973 004 For your notes...
Страница 471: ...13 BIS Module E BIS Rev 01 Technical Manual...
Страница 472: ......
Страница 480: ...Module Frames and Modules 13 6 2062973 004 2 Configuration There is no configuration for E BIS module...
Страница 494: ...Module Frames and Modules 13 20 2062973 004 For your notes...
Страница 495: ...14 NeuroMuscular Transmission Module E NMT Rev 01 Technical Manual...
Страница 496: ......
Страница 504: ...Module Frames and Modules 14 6 2062973 004 2 Configuration There is no configuration for the E NMT module...
Страница 518: ...Module Frames and Modules 14 20 2062973 004 For your notes...
Страница 520: ...Module Frames and Modules 14 22 2062973 004 For your notes...
Страница 521: ...15 F5 5 Module Frame F7 7 Module Frame Technical Manual...
Страница 522: ......
Страница 532: ...Module Frames and Modules 15 8 2062973 004 2 Configuration There is no configuration for the F5 and F7 module frames...
Страница 558: ...Module Frames and Modules 15 34 2062973 004 For your notes...
Страница 560: ...Module Frames and Modules 15 36 2062973 004 For your notes...
Страница 561: ...16 Patient Data Module v2 Technical Manual...
Страница 562: ......
Страница 620: ...Module Frames and Modules 16 56 2062973 004...
Страница 621: ......