![GE HEALTHCARE E-PSMP Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/ge-healthcare/e-psmp/e-psmp_technical-manual_3949047024.webp)
Module Frames and Modules
2 - 10
2062973-004
ECG amplifiers and baseline restoration
The function of the ECG amplifiers and baseline restoration is to amplify the signal and to
restore the baseline of the signal in the middle of the display after the change of the signal
level, e.g. after the change of the DC offset voltage.
Pacer detection
Pacer detection has been made by using four slew rate detector circuits. The pacer detection
amplifiers have been realized at the front of the slew rate detectors independently of the ECG
measuring channels.
Respiration impedance supply
The 31.25 kHz sine wave generator is used as the respiration measurement signal supply.
Analog switches are used for connecting the sine wave to the ECG leads to be measured.
Respiration impedance amplifiers
Buffer amplifiers are used in respiration measurement. Analog switches are used for selecting
the measurement leads. There are also additional amplifiers for increasing the respiration
signal gain. When ECG measurement is 5/12-lead5/12-lead, the respiration measurement is
always done between R and F, independently on the ECG lead selection. When ECG
measurement is 3-lead, then the respiration measurement is happened at the same lead as
the ECG measurement (I, II, or III).
ECG CPU
The CPU is a 16 bit H8/3052 single-chip microcomputer. It contains 128 kbytes of flash memory
and 4 kbytes of RAM. The clock frequency is 16 MHz.
RS485 communication
The communication to the CPU board of the monitor uses RS485 protocol. The RS485 driver
circuits are optically isolated from the processor of the module.
Power supply
The ECG board has a driver-controlled half-bridge switching power supply with 5 kV isolation.
The supply voltages have been regulated with linear regulators.
ECG filtering
There are three ECG filtering modes:
MONITORING
0.5 to 30 Hz (with 50 Hz reject filter)
0.5 to 40 Hz (with 60 Hz reject filter)
DIAGNOSTIC 12-lead ECG
0.05 to 150 Hz
ST FILTER
0.05 to 30 Hz (with 50 Hz reject filter)
0.05 to 40 Hz (with 60 Hz reject filter)
The purpose of filtering is to reduce high frequency noise and low frequency (e.g. respiratory)
movement artifacts.
The monitor filter is used in normal monitoring. The diagnostic filter is used if more accurate
diagnostic information is needed. The ST filter gives more accurate information of ST segment,
but reduces high frequency noise.
The high-pass filters 0.5 Hz and 0.05 Hz are done with software. The monitor sends a command
to the hemodynamic module determining which of the corner frequencies 0.5 Hz or 0.05 Hz is
to be used.
The 50 Hz and 60 Hz reject filters are both low-pass filters with zero at 50 Hz or 60 Hz
correspondingly. They are software based filters used for the mains supply filtering. With these
filters the 3 dB value for low-pass filter is 30 Hz or 40 Hz.
Содержание E-PSMP
Страница 2: ......
Страница 4: ...Module Frames and Modules 2062973 004 For your notes...
Страница 6: ...Module Frames and Modules 1 ii 2062973 004...
Страница 10: ...Module Frames and Modules 1 4 2062973 004 For your notes...
Страница 11: ...2 Hemodynamic Modules E PRESTN E RESTN E PRETN Rev 00 E PSMP E PSM Rev 01 Technical Manual...
Страница 12: ......
Страница 92: ...Module Frames and Modules 2 78 2062973 004 For your notes...
Страница 94: ...Module Frames and Modules 2 80 2062973 004 For your notes...
Страница 95: ...3 Continuous Cardiac Output Module E PiCCO Technical Manual...
Страница 96: ......
Страница 124: ...Module Frames and Modules 3 26 2062973 004 For your notes...
Страница 125: ...4 Cardiac Output and SvO2 Module E COPSv Rev 01 Cardiac Output Module E COP Rev 01 Technical Manual...
Страница 126: ......
Страница 138: ...Module Frames and Modules 4 10 2062973 004 2 Configuration There is no configuration for the E COP or E COPSv modules...
Страница 159: ...Cardiac Output Modules E COPSv E COP 4 31 2062973 004...
Страница 160: ...Module Frames and Modules 4 32 2062973 004 For your notes...
Страница 162: ...Module Frames and Modules 4 34 2062973 004 For your notes...
Страница 163: ...5 Dual Pressure Module E PP Pressure Temp Module E PT Pressure Module E P Technical Manual...
Страница 164: ......
Страница 198: ...Module Frames and Modules 5 32 2062973 004 For your notes...
Страница 199: ...6 Masimo Module E MASIMO Technical Manual...
Страница 200: ......
Страница 202: ...Module Frames and Modules 6 ii 2062973 004...
Страница 207: ...Masimo Module E MASIMO 6 5 2062973 004 2 Configuration There is no configuration for the E MASIMO module...
Страница 218: ...Module Frames and Modules 6 16 2062973 004 For your notes...
Страница 220: ...Module Frames and Modules 6 18 2062973 004 For your notes...
Страница 221: ...7 Nellcor Compatible Module E NSAT E NSATX Technical Manual...
Страница 222: ......
Страница 224: ...Module Frames and Modules 7 ii 2062973 004...
Страница 236: ...Module Frames and Modules 7 12 2062973 004...
Страница 244: ...Module Frames and Modules 7 20 2062973 004 For your notes...
Страница 246: ...Module Frames and Modules 7 22 2062973 004 For your notes...
Страница 247: ...8 Respiratory Modules E sCAiOVX E sCAiOV E sCAiO E sCOVX E sCOV E sCO Technical Manual...
Страница 248: ......
Страница 284: ...Module Frames and Modules 8 34 2062973 004...
Страница 316: ...Module Frames and Modules 8 66 2062973 004 For your notes...
Страница 319: ...9 Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV E CO Technical Manual...
Страница 320: ......
Страница 336: ...Module Frames and Modules 9 14 2062973 004 For your notes...
Страница 353: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 31 2062973 004...
Страница 379: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 57 2062973 004...
Страница 380: ...Module Frames and Modules 9 58 2062973 004 For your notes...
Страница 381: ...10 Single width Airway Module E miniC Technical Manual...
Страница 382: ......
Страница 391: ...Single width Airway Module E miniC 10 9 2062973 004 2 Configuration There is no configuration for the E miniC module...
Страница 397: ...Single width Airway Module E miniC 10 15 2062973 004 Complete on the Maintenance check form on page 10 36...
Страница 410: ...Module Frames and Modules 10 28 2062973 004...
Страница 419: ...11 Entropy Module E ENTROPY Rev 01 Technical Manual...
Страница 420: ......
Страница 422: ...Module Frames and Modules 11 ii 2062973 004 7 2 Entropy Module E ENTROPY 01 11 17 Maintenance check form 11 19...
Страница 427: ...Entropy Module E ENTROPY Rev 01 11 5 2062973 004 2 Configuration There is no configuration for the E ENTROPY module...
Страница 442: ...Module Frames and Modules 11 20 2062973 004 For your notes...
Страница 443: ...12 EEG Module E EEG and EEG Headbox N EEG Technical Manual...
Страница 444: ......
Страница 446: ...Module Frames and Modules 12 ii 2062973 004 7 3 EEG Headbox N EEG 12 20 Maintenance check form 12 23...
Страница 452: ...Module Frames and Modules 12 6 2062973 004 2 Configuration There is no configuration for the E EEG module...
Страница 468: ...Module Frames and Modules 12 22 2062973 004 For your notes...
Страница 470: ...Module Frames and Modules 12 24 2062973 004 For your notes...
Страница 471: ...13 BIS Module E BIS Rev 01 Technical Manual...
Страница 472: ......
Страница 480: ...Module Frames and Modules 13 6 2062973 004 2 Configuration There is no configuration for E BIS module...
Страница 494: ...Module Frames and Modules 13 20 2062973 004 For your notes...
Страница 495: ...14 NeuroMuscular Transmission Module E NMT Rev 01 Technical Manual...
Страница 496: ......
Страница 504: ...Module Frames and Modules 14 6 2062973 004 2 Configuration There is no configuration for the E NMT module...
Страница 518: ...Module Frames and Modules 14 20 2062973 004 For your notes...
Страница 520: ...Module Frames and Modules 14 22 2062973 004 For your notes...
Страница 521: ...15 F5 5 Module Frame F7 7 Module Frame Technical Manual...
Страница 522: ......
Страница 532: ...Module Frames and Modules 15 8 2062973 004 2 Configuration There is no configuration for the F5 and F7 module frames...
Страница 558: ...Module Frames and Modules 15 34 2062973 004 For your notes...
Страница 560: ...Module Frames and Modules 15 36 2062973 004 For your notes...
Страница 561: ...16 Patient Data Module v2 Technical Manual...
Страница 562: ......
Страница 620: ...Module Frames and Modules 16 56 2062973 004...
Страница 621: ......