Respiratory Modules, E-sCAiOVX, E-sCAiOV, E-sCAiO, E-sCOVX, E-sCOV, E-sCO
8 - 49
2062973-004
measuring unit itself, the entire module should be replaced and the faulty module be sent to GE
for repair.
NOTE: The MiniTPX measuring unit can only be repaired and calibrated at the factory. In case of
failure, the entire module should be replaced and the faulty module be sent to GE for repair.
6.1.3 ESD precautions
All external connectors of the module are designed with protection from ESD damage.
However, if the module requires service, exposed components and assemblies inside are
susceptible to ESD damage. This includes human hands, non-ESD protected work stations or
improperly grounded test equipment. The following guidelines may not guarantee a 100%
static-free workstation, but can greatly reduce the potential for failure of any electronic
assemblies being serviced:
•
Discharge any static charge you may have built up before handling semiconductors or
assemblies containing semiconductors.
•
A grounded, antistatic wristband or heel strap should be worn at all times while handling
or repairing assemblies containing semiconductors.
•
Use properly grounded test equipment.
•
Use a static-free work surface while handling or working on assemblies containing
semiconductors.
•
Do not remove semiconductors or assemblies containing semiconductors from antistatic
containers until absolutely necessary.
•
Do not slide semiconductors or electrical/electronic assemblies across any surface.
•
Do not touch semiconductor leads unless absolutely necessary.
•
Semiconductors and electronic assemblies should be stored only in antistatic bags or
boxes.
•
Handle all PCB assemblies by their edges.
•
Do not flex or twist a circuit board.
6.1.4 Protection from dust
The gas sampling system must be protected from dust entering the tubes, valves and other
components. In order to achieve this goal, the following measures must be taken:
•
Have the D-fend Pro water trap always connected to the module.
•
Clean and dust free working environment during all service procedures.
•
Minimize the times with any open connections in the gas sampling system.
•
Always close the open tube connections of the sampling system when not working on the
module.
WARNING
Protect module from electrostatic discharge.
WARNING
Module must be handled with care to prevent dust
from entering the gas sampling system.
Содержание E-PSMP
Страница 2: ......
Страница 4: ...Module Frames and Modules 2062973 004 For your notes...
Страница 6: ...Module Frames and Modules 1 ii 2062973 004...
Страница 10: ...Module Frames and Modules 1 4 2062973 004 For your notes...
Страница 11: ...2 Hemodynamic Modules E PRESTN E RESTN E PRETN Rev 00 E PSMP E PSM Rev 01 Technical Manual...
Страница 12: ......
Страница 92: ...Module Frames and Modules 2 78 2062973 004 For your notes...
Страница 94: ...Module Frames and Modules 2 80 2062973 004 For your notes...
Страница 95: ...3 Continuous Cardiac Output Module E PiCCO Technical Manual...
Страница 96: ......
Страница 124: ...Module Frames and Modules 3 26 2062973 004 For your notes...
Страница 125: ...4 Cardiac Output and SvO2 Module E COPSv Rev 01 Cardiac Output Module E COP Rev 01 Technical Manual...
Страница 126: ......
Страница 138: ...Module Frames and Modules 4 10 2062973 004 2 Configuration There is no configuration for the E COP or E COPSv modules...
Страница 159: ...Cardiac Output Modules E COPSv E COP 4 31 2062973 004...
Страница 160: ...Module Frames and Modules 4 32 2062973 004 For your notes...
Страница 162: ...Module Frames and Modules 4 34 2062973 004 For your notes...
Страница 163: ...5 Dual Pressure Module E PP Pressure Temp Module E PT Pressure Module E P Technical Manual...
Страница 164: ......
Страница 198: ...Module Frames and Modules 5 32 2062973 004 For your notes...
Страница 199: ...6 Masimo Module E MASIMO Technical Manual...
Страница 200: ......
Страница 202: ...Module Frames and Modules 6 ii 2062973 004...
Страница 207: ...Masimo Module E MASIMO 6 5 2062973 004 2 Configuration There is no configuration for the E MASIMO module...
Страница 218: ...Module Frames and Modules 6 16 2062973 004 For your notes...
Страница 220: ...Module Frames and Modules 6 18 2062973 004 For your notes...
Страница 221: ...7 Nellcor Compatible Module E NSAT E NSATX Technical Manual...
Страница 222: ......
Страница 224: ...Module Frames and Modules 7 ii 2062973 004...
Страница 236: ...Module Frames and Modules 7 12 2062973 004...
Страница 244: ...Module Frames and Modules 7 20 2062973 004 For your notes...
Страница 246: ...Module Frames and Modules 7 22 2062973 004 For your notes...
Страница 247: ...8 Respiratory Modules E sCAiOVX E sCAiOV E sCAiO E sCOVX E sCOV E sCO Technical Manual...
Страница 248: ......
Страница 284: ...Module Frames and Modules 8 34 2062973 004...
Страница 316: ...Module Frames and Modules 8 66 2062973 004 For your notes...
Страница 319: ...9 Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV E CO Technical Manual...
Страница 320: ......
Страница 336: ...Module Frames and Modules 9 14 2062973 004 For your notes...
Страница 353: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 31 2062973 004...
Страница 379: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 57 2062973 004...
Страница 380: ...Module Frames and Modules 9 58 2062973 004 For your notes...
Страница 381: ...10 Single width Airway Module E miniC Technical Manual...
Страница 382: ......
Страница 391: ...Single width Airway Module E miniC 10 9 2062973 004 2 Configuration There is no configuration for the E miniC module...
Страница 397: ...Single width Airway Module E miniC 10 15 2062973 004 Complete on the Maintenance check form on page 10 36...
Страница 410: ...Module Frames and Modules 10 28 2062973 004...
Страница 419: ...11 Entropy Module E ENTROPY Rev 01 Technical Manual...
Страница 420: ......
Страница 422: ...Module Frames and Modules 11 ii 2062973 004 7 2 Entropy Module E ENTROPY 01 11 17 Maintenance check form 11 19...
Страница 427: ...Entropy Module E ENTROPY Rev 01 11 5 2062973 004 2 Configuration There is no configuration for the E ENTROPY module...
Страница 442: ...Module Frames and Modules 11 20 2062973 004 For your notes...
Страница 443: ...12 EEG Module E EEG and EEG Headbox N EEG Technical Manual...
Страница 444: ......
Страница 446: ...Module Frames and Modules 12 ii 2062973 004 7 3 EEG Headbox N EEG 12 20 Maintenance check form 12 23...
Страница 452: ...Module Frames and Modules 12 6 2062973 004 2 Configuration There is no configuration for the E EEG module...
Страница 468: ...Module Frames and Modules 12 22 2062973 004 For your notes...
Страница 470: ...Module Frames and Modules 12 24 2062973 004 For your notes...
Страница 471: ...13 BIS Module E BIS Rev 01 Technical Manual...
Страница 472: ......
Страница 480: ...Module Frames and Modules 13 6 2062973 004 2 Configuration There is no configuration for E BIS module...
Страница 494: ...Module Frames and Modules 13 20 2062973 004 For your notes...
Страница 495: ...14 NeuroMuscular Transmission Module E NMT Rev 01 Technical Manual...
Страница 496: ......
Страница 504: ...Module Frames and Modules 14 6 2062973 004 2 Configuration There is no configuration for the E NMT module...
Страница 518: ...Module Frames and Modules 14 20 2062973 004 For your notes...
Страница 520: ...Module Frames and Modules 14 22 2062973 004 For your notes...
Страница 521: ...15 F5 5 Module Frame F7 7 Module Frame Technical Manual...
Страница 522: ......
Страница 532: ...Module Frames and Modules 15 8 2062973 004 2 Configuration There is no configuration for the F5 and F7 module frames...
Страница 558: ...Module Frames and Modules 15 34 2062973 004 For your notes...
Страница 560: ...Module Frames and Modules 15 36 2062973 004 For your notes...
Страница 561: ...16 Patient Data Module v2 Technical Manual...
Страница 562: ......
Страница 620: ...Module Frames and Modules 16 56 2062973 004...
Страница 621: ......