Cleaning and care
9-3
Regular checks
If the monitor does not work as described, see section
When you start monitoring, check that:
•
The module is firmly in place.
•
Accessories are intact and properly connected.
•
You have selected desired parameters to be displayed in the digit and waveform fields.
ECG and impedance respiration
•
After connecting ECG cable, check that the message 'Leads off' display in waveform field .
NIBP
•
After connecting NIBP hose, check that the message ‘Adult/Pediatric’ or ‘Neonatal’ will
display in NIBP digital field for several seconds.
Pulse oximetry
•
After connecting SpO2 cable and sensor, check that the red light is lit in the sensor .
Temperature
•
After connecting Temp cable and sensor, check that the message ‘Performing temp test:’
will display in Temperature digital field for several seconds.
IBP
•
After connecting IBP cable and transducer, check that the message ‘InvBP’s not Zeroed’
will display in message field.
Airway gas
•
Make sure the acquisition module is connected to the monitor.
•
After installing the module, ‘Calibrating gas sensor’ will display in gas waveform field for
about 1 minutes.
Entropy
•
Make sure the acquisition module is connected to the monitor.
•
Connect the Entropy sensor cable to the module, and check that the “No sensor”
message apprears.
Functioning of the alarms
•
Check that the red, yellow and cyan alarm lights are lit and the speaker gives an audible
beep momentarily when the monitor starts up.
Safety checks for software
The GE Healthcare software design controls include performance of a risk analysis using
methods consistent with ISO 14971 Medical devices - Application of risk management to
medical devices.
The monitor software employs watchdog timers, self-monitoring activities (memory,
communication and sensor checks and so on), and power-on self-diagnostics (for example,
memory checksums).
Содержание B20
Страница 2: ......
Страница 16: ...B40 B20 Patient Monitor xii ...
Страница 17: ...1 Introduction ...
Страница 18: ......
Страница 27: ...2 System description ...
Страница 28: ......
Страница 58: ...B40 B20 Patient Monitor 2 30 ...
Страница 59: ...3 Installation ...
Страница 60: ......
Страница 69: ...4 Monitoring basic ...
Страница 70: ......
Страница 79: ...5 Alarms ...
Страница 80: ......
Страница 102: ...B40 B20 Patient Monitor 5 22 ...
Страница 103: ...6 Monitor setup ...
Страница 104: ......
Страница 123: ...7 Trends ...
Страница 124: ......
Страница 140: ...B40 B20 Patient Monitor 7 16 ...
Страница 141: ...8 Print and record ...
Страница 142: ......
Страница 153: ...9 Cleaning and care ...
Страница 154: ......
Страница 163: ...10 Troubleshooting ...
Страница 164: ......
Страница 181: ...11 ECG ...
Страница 182: ......
Страница 205: ...12 Impedance respiration ...
Страница 206: ......
Страница 215: ...13 Pulse oximetry ...
Страница 216: ......
Страница 232: ...B40 B20 Patient Monitor 13 16 ...
Страница 233: ...14 Non invasive blood pressure ...
Страница 234: ......
Страница 248: ...B40 B20 Patient Monitor 14 14 ...
Страница 249: ...15 Invasive blood pressure ...
Страница 250: ......
Страница 261: ...16 Temperature ...
Страница 262: ......
Страница 267: ...17 Airway gas ...
Страница 268: ......
Страница 293: ...18 Entropy ...
Страница 294: ......
Страница 316: ...Abbreviations A 12 ...
Страница 318: ...B40 B20 Patient Monitor B 2 ...
Страница 319: ......