Troubleshooting
10-1
10 Troubleshooting
Overview
The troubleshooting section consists of two parts which together should help you resolve the
most common monitoring problems. The parts are
.
Checklist
Check the following things before monitoring to ensure that you have remembered to make all
essential preparations, and if any problems occur during monitoring. Parameter-specific
checklists can be found in the respective parameter sections.
Check that:
•
The monitor and its module do not have any visual defects such as cracks or loose parts.
•
The batteries are inserted and charged.
•
The power cord is connected to an electrical wall outlet and to the monitor.
•
The module are pushed properly into the frame and the monitor recognizes the module
(parameter is displayed on the screen and menu selections are active).
•
Patient connection cables are attached to the module connectors.
•
Trends of the previous patient are erased.
•
Alarm limits are suitable for the patient.
•
The desired measurements have been selected for screen through
Monitor Setup
–
Screen Setup - Waveform Fields
or
Digit Fields
.
•
Disposable accessories are not reused.
•
Sensors are not expired.
Power interruption
If the monitor is turned off, trend data and the latest user-made settings remain in the monitor
memory for 15 minutes even if the power (mains) is interrupted. After 15 minutes, trend data is
lost and the monitor returns to the user default settings (start-up mode).
Содержание B20
Страница 2: ......
Страница 16: ...B40 B20 Patient Monitor xii ...
Страница 17: ...1 Introduction ...
Страница 18: ......
Страница 27: ...2 System description ...
Страница 28: ......
Страница 58: ...B40 B20 Patient Monitor 2 30 ...
Страница 59: ...3 Installation ...
Страница 60: ......
Страница 69: ...4 Monitoring basic ...
Страница 70: ......
Страница 79: ...5 Alarms ...
Страница 80: ......
Страница 102: ...B40 B20 Patient Monitor 5 22 ...
Страница 103: ...6 Monitor setup ...
Страница 104: ......
Страница 123: ...7 Trends ...
Страница 124: ......
Страница 140: ...B40 B20 Patient Monitor 7 16 ...
Страница 141: ...8 Print and record ...
Страница 142: ......
Страница 153: ...9 Cleaning and care ...
Страница 154: ......
Страница 163: ...10 Troubleshooting ...
Страница 164: ......
Страница 181: ...11 ECG ...
Страница 182: ......
Страница 205: ...12 Impedance respiration ...
Страница 206: ......
Страница 215: ...13 Pulse oximetry ...
Страница 216: ......
Страница 232: ...B40 B20 Patient Monitor 13 16 ...
Страница 233: ...14 Non invasive blood pressure ...
Страница 234: ......
Страница 248: ...B40 B20 Patient Monitor 14 14 ...
Страница 249: ...15 Invasive blood pressure ...
Страница 250: ......
Страница 261: ...16 Temperature ...
Страница 262: ......
Страница 267: ...17 Airway gas ...
Страница 268: ......
Страница 293: ...18 Entropy ...
Страница 294: ......
Страница 316: ...Abbreviations A 12 ...
Страница 318: ...B40 B20 Patient Monitor B 2 ...
Страница 319: ......