System description
2-27
Anesthetic agents specified
measurement range
Hal, Enf, Iso: 0 to 6 vol%
Sev: 0 to 8 vol%
Des: 0 to 20 vol%
Anesthetic agents
measurement accuracy
± (0.15 vol% + 5% of reading)
Anesthetic agents display
resolution
0.01% for values between 0…1%
0.1% for values equal or above 1%
Anesthetic agents total system
response time
< 3.5 seconds with a 2 or 3m sampling line (< 3.8 for
Hal)
Anesthetic agents rise time
Hal, Enf, Iso, Des: < 420 ms with a 3 m sampling line
Hal: < 800 ms with a 3 m sampling line
Hal, Enf, Iso, Des: < 700 ms with a 6 m sampling line
Hal: < 1800 ms with a 6 m sampling line
Hal drift
< 0.1 vol%
Enf drift
< 0.1 vol%
Iso drift
< 0.1 vol%
Sev drift
< 0.1 vol%
Des drift
< 0.3 vol%
The E-sCAiO, N-CAiO modules automatically identify the anesthetic agent present in
the sampled gas and measure the concentration of the identified agent.
Identification threshold
0.15 vol%
Identification time
< 20 s
The E-sCAiO module automatically identifies mixtures of two anesthetic agents present
in the sampled gas and measures the concentrations of the two identified agents.
Identification threshold for the
second agent at 1 MAC of the
first agent:
0.2 vol% +10% of the concentration of the first agent
Effects of interfering gases and vapors
E-miniC module
Non-disturbing gases of
which effect on CO
2
(5 vol%)
readings < 0.2 vol%:
- Ethanol C
2
H
5
OH (<0.3%)
- Acetone (<0.1% )
- Methane CH
4
(<0.2%)
- Nitrogen N
2
(0-100%)
- Water vapor (0-100%)
- Trichloromonofluoromethane (<1%)
- Dichlorotetrafluoroethane (<1%)
- Dichlorofluromethane (<1%)
Содержание B20
Страница 2: ......
Страница 16: ...B40 B20 Patient Monitor xii ...
Страница 17: ...1 Introduction ...
Страница 18: ......
Страница 27: ...2 System description ...
Страница 28: ......
Страница 58: ...B40 B20 Patient Monitor 2 30 ...
Страница 59: ...3 Installation ...
Страница 60: ......
Страница 69: ...4 Monitoring basic ...
Страница 70: ......
Страница 79: ...5 Alarms ...
Страница 80: ......
Страница 102: ...B40 B20 Patient Monitor 5 22 ...
Страница 103: ...6 Monitor setup ...
Страница 104: ......
Страница 123: ...7 Trends ...
Страница 124: ......
Страница 140: ...B40 B20 Patient Monitor 7 16 ...
Страница 141: ...8 Print and record ...
Страница 142: ......
Страница 153: ...9 Cleaning and care ...
Страница 154: ......
Страница 163: ...10 Troubleshooting ...
Страница 164: ......
Страница 181: ...11 ECG ...
Страница 182: ......
Страница 205: ...12 Impedance respiration ...
Страница 206: ......
Страница 215: ...13 Pulse oximetry ...
Страница 216: ......
Страница 232: ...B40 B20 Patient Monitor 13 16 ...
Страница 233: ...14 Non invasive blood pressure ...
Страница 234: ......
Страница 248: ...B40 B20 Patient Monitor 14 14 ...
Страница 249: ...15 Invasive blood pressure ...
Страница 250: ......
Страница 261: ...16 Temperature ...
Страница 262: ......
Страница 267: ...17 Airway gas ...
Страница 268: ......
Страница 293: ...18 Entropy ...
Страница 294: ......
Страница 316: ...Abbreviations A 12 ...
Страница 318: ...B40 B20 Patient Monitor B 2 ...
Страница 319: ......