B40/B20 Patient Monitor
10-6
NIBP call service error
NIBP error
•
Contact authorized service
personnel.
NIBP manual
Autocycling mode is interrupted
because of an air leak or loose cuff.
•
Check the NIBP setup and restart
autocycling.
NIBP cuff loose
Cuff is not attached to the patient or it
is too loose. The hose is not connected
to the module.
•
Check cuff and hose.
NIBP cuff occlusion
Tubes or hose are kinked.
•
Check tubes and hose.
NIBP cuff overpressure
Cuff is squeezed during measurement
and pressure safety limits are
exceeded.
•
Check cuff, hose and tubes.
•
Restart measurement.
No battery backup
The monitor has the battery option but
there is no battery inserted.
•
Insert a battery.
No Entropy sensor
The sensor is not connected to the
cable.
•
Check connection between Entropy
sensor and cable.
No xx transducer
Transducer or channel x cable
disconnected.
•
Connect the transducer or the
cable.
No SpO2 probe
There is no SpO
2
probe.
•
Check connection between the
SpO
2
sensor and module.
No SpO2 pulse
Pulse signal is poor.
•
Try other measurement sites.
No printer selected
There is no printer selected on the
monitor.
•
Select a printer, see section
chapter.
Patient discharged
The patient has just been discharged.
•
For more information, see section
chapter.
Patient admitted
The current patient has just been
admitted.
•
For more information, see section
chapter.
P1 over range
P2 over range
Measured pressure is out of alarm
limit’s max value
•
Check the patient status.
P1 under range
P2 under range
Measured pressure is out of alarm
limit’s min value
•
Check the patient status.
Printer error
Printer is not working properly.
•
Check that the network printer is
operational.
Printing...
Printing network printer has started.
•
Please wait until the printing is
finished.
Printing ready
Monitor has completed sending
printing data to the printer.
•
Please wait until the printing is
finished.
Recorder: out of paper
The recorder is out of paper.
•
Add paper.
Recorder: cover open
The recorder cover is open.
•
Close the recorder cover.
Message
Explanation
What to do
Содержание B20
Страница 2: ......
Страница 16: ...B40 B20 Patient Monitor xii ...
Страница 17: ...1 Introduction ...
Страница 18: ......
Страница 27: ...2 System description ...
Страница 28: ......
Страница 58: ...B40 B20 Patient Monitor 2 30 ...
Страница 59: ...3 Installation ...
Страница 60: ......
Страница 69: ...4 Monitoring basic ...
Страница 70: ......
Страница 79: ...5 Alarms ...
Страница 80: ......
Страница 102: ...B40 B20 Patient Monitor 5 22 ...
Страница 103: ...6 Monitor setup ...
Страница 104: ......
Страница 123: ...7 Trends ...
Страница 124: ......
Страница 140: ...B40 B20 Patient Monitor 7 16 ...
Страница 141: ...8 Print and record ...
Страница 142: ......
Страница 153: ...9 Cleaning and care ...
Страница 154: ......
Страница 163: ...10 Troubleshooting ...
Страница 164: ......
Страница 181: ...11 ECG ...
Страница 182: ......
Страница 205: ...12 Impedance respiration ...
Страница 206: ......
Страница 215: ...13 Pulse oximetry ...
Страница 216: ......
Страница 232: ...B40 B20 Patient Monitor 13 16 ...
Страница 233: ...14 Non invasive blood pressure ...
Страница 234: ......
Страница 248: ...B40 B20 Patient Monitor 14 14 ...
Страница 249: ...15 Invasive blood pressure ...
Страница 250: ......
Страница 261: ...16 Temperature ...
Страница 262: ......
Страница 267: ...17 Airway gas ...
Страница 268: ......
Страница 293: ...18 Entropy ...
Страница 294: ......
Страница 316: ...Abbreviations A 12 ...
Страница 318: ...B40 B20 Patient Monitor B 2 ...
Страница 319: ......