B40/B20 Patient Monitor
6-16
Configuring trends and snapshot
Changing default trend
You can select graphical or numerical trends to be displayed by default:
1.
Press the
Monitor Setup
key.
2.
Select
Install/Service
and enter the password.
3.
Select
Trends & Snapshot - Default Trend
Configuring trend pages
You can change the parameters on the trend fields:
1.
Press the
Monitor Setup
key.
2.
Select
Install/Service
and enter the password.
3.
Select
Trends & Snapshot - Graphical Trends
.
4.
Select the trend page that you want to change.
5.
Select graphical parameters for each field.
The field numbers start from the top of the screen. Select one parameter for each field on the
trend page, or turn the field
OFF
. When all the fields are
OFF
, the page is displayed with empty
fields. The time scale and page number appear at the bottom of the page.
Field 1-5 can be displayed on screen and Field 6 can be printed.
If several similar fields are selected on top of each other, they form one higher field. Equal fields
cannot be defined separate to each other.
NOTE: You cannot make changes in numerical trend page configuration.
Trends & Snapshot
Default Trend
Previous Menu
Graph
Change default trend type:
graphical or numerical.
Install/Service
Graphical Trends
Snapshot
Monitor
Setup
Содержание B20
Страница 2: ......
Страница 16: ...B40 B20 Patient Monitor xii ...
Страница 17: ...1 Introduction ...
Страница 18: ......
Страница 27: ...2 System description ...
Страница 28: ......
Страница 58: ...B40 B20 Patient Monitor 2 30 ...
Страница 59: ...3 Installation ...
Страница 60: ......
Страница 69: ...4 Monitoring basic ...
Страница 70: ......
Страница 79: ...5 Alarms ...
Страница 80: ......
Страница 102: ...B40 B20 Patient Monitor 5 22 ...
Страница 103: ...6 Monitor setup ...
Страница 104: ......
Страница 123: ...7 Trends ...
Страница 124: ......
Страница 140: ...B40 B20 Patient Monitor 7 16 ...
Страница 141: ...8 Print and record ...
Страница 142: ......
Страница 153: ...9 Cleaning and care ...
Страница 154: ......
Страница 163: ...10 Troubleshooting ...
Страница 164: ......
Страница 181: ...11 ECG ...
Страница 182: ......
Страница 205: ...12 Impedance respiration ...
Страница 206: ......
Страница 215: ...13 Pulse oximetry ...
Страница 216: ......
Страница 232: ...B40 B20 Patient Monitor 13 16 ...
Страница 233: ...14 Non invasive blood pressure ...
Страница 234: ......
Страница 248: ...B40 B20 Patient Monitor 14 14 ...
Страница 249: ...15 Invasive blood pressure ...
Страница 250: ......
Страница 261: ...16 Temperature ...
Страница 262: ......
Страница 267: ...17 Airway gas ...
Страница 268: ......
Страница 293: ...18 Entropy ...
Страница 294: ......
Страница 316: ...Abbreviations A 12 ...
Страница 318: ...B40 B20 Patient Monitor B 2 ...
Страница 319: ......