Airway gas
17-13
Calibrating
Perform calibration every six months in normal use, once every two months in continuous use,
and whenever there are indications of errors in the gas readings to ensure that the
measurement accuracy remains within specifications.
Calibration setup
Required tools
•
P/N: 755534-HEL Calibration Gas Regulator
•
P/N: M1006864, Calibration Gas Regulator, US only
•
3 m / 10 ft anesthesia gas sampling line
•
For E-sCO, E-sCAiO, N-CAiO module:
−
P/N: 755583-HEL Calibration gas, CO2, O2, N2O, DESF, package of 1 can (with E-
sCAiO or N-CAiO module)
−
P/N: 755581-HEL QUICK CAL calibration gas, CO2, O2, N2O, package of 4 cans (with
E-sCO module)
−
P/N: 755571-HEL, Calibration Gas, 5% CO2, 54.5% O2, 36.0% N2O, 2.0%
DESFLURANE, BAL N2 (with E-sCAiO or N-CAiO module) US only
−
P/N: 755587, Calibration Gas, CO2, O2, Balance, 4 cans/pkg (with E-sCO module) US
only
NOTE: Use only the specified GE Healthcare calibration gas for the gas calibration to
ensure measurement accuracy. Do not use any other calibration gases. Check the
calibration gas container's labelling to ensure that the calibration gas has not expired.
NOTE: Ensure that the gas regulator is functioning properly before gas calibration.
Refer to the gas regulator's "Instructions for Use" letter for the annual maintenance
instructions.
Connections
1.
Ensure that the module is connected to the monitor.
2.
Ensure that you have a new D-fend or D-fend Pro water trap in use.
3.
Connect the gas regulator to the calibration gas container.
Figure 17-5 Attaching regulator to calibration can
4.
Connect a new gas sampling line to the sampling line connector in the water trap.
5.
Connect the other end of the gas sampling line to the regulator on the gas container.
Leave the regulator overflow port open to room air.
Содержание B20
Страница 2: ......
Страница 16: ...B40 B20 Patient Monitor xii ...
Страница 17: ...1 Introduction ...
Страница 18: ......
Страница 27: ...2 System description ...
Страница 28: ......
Страница 58: ...B40 B20 Patient Monitor 2 30 ...
Страница 59: ...3 Installation ...
Страница 60: ......
Страница 69: ...4 Monitoring basic ...
Страница 70: ......
Страница 79: ...5 Alarms ...
Страница 80: ......
Страница 102: ...B40 B20 Patient Monitor 5 22 ...
Страница 103: ...6 Monitor setup ...
Страница 104: ......
Страница 123: ...7 Trends ...
Страница 124: ......
Страница 140: ...B40 B20 Patient Monitor 7 16 ...
Страница 141: ...8 Print and record ...
Страница 142: ......
Страница 153: ...9 Cleaning and care ...
Страница 154: ......
Страница 163: ...10 Troubleshooting ...
Страница 164: ......
Страница 181: ...11 ECG ...
Страница 182: ......
Страница 205: ...12 Impedance respiration ...
Страница 206: ......
Страница 215: ...13 Pulse oximetry ...
Страница 216: ......
Страница 232: ...B40 B20 Patient Monitor 13 16 ...
Страница 233: ...14 Non invasive blood pressure ...
Страница 234: ......
Страница 248: ...B40 B20 Patient Monitor 14 14 ...
Страница 249: ...15 Invasive blood pressure ...
Страница 250: ......
Страница 261: ...16 Temperature ...
Страница 262: ......
Страница 267: ...17 Airway gas ...
Страница 268: ......
Страница 293: ...18 Entropy ...
Страница 294: ......
Страница 316: ...Abbreviations A 12 ...
Страница 318: ...B40 B20 Patient Monitor B 2 ...
Страница 319: ......