B40/B20 Patient Monitor
14-2
•
Periodically check patient limb circulation distal to the cuff. Check frequently when using
Auto NIBP in one and two minute intervals. The one and two minute intervals are not
recommended for extended periods of time.
•
The monitor sets the inflation pressure automatically according to the previous
measurement. Reset the case or discharge the patient to reset the inflation limits before
measuring NIBP on a new patient.
Overview
The non-invasive blood pressure (NIBP) measurement uses the oscillometric measuring
principle. The cuff is inflated with a pressure slightly higher than the presumed systolic
pressure, then slowly deflated at a speed based on the patient’s heart rate, collecting data
from the oscillations produced by the pulsating artery. Based on this data, the unit calculates
values for systolic, mean and diastolic pressures.
Blood pressure measurements determined with this device are equivalent to those obtained by
an intra-arterial blood pressure measurement device. Multiple intra-arterial sites were used.
You can set the NIBP module into an automatic cycling mode to make measurements at
desired time intervals. You can also measure NIBP continuously for five minutes in STAT mode
or take separate single measurements.
NOTE: Intervals below 10 minutes and STAT measurements are not recommended for
extended periods of time.
NOTE: This equipment is suitable for use in the presence of electrosurgery.
Figure 14-1
N
on-invasive blood pressure measurement connector
NIBP
connector
Содержание B20
Страница 2: ......
Страница 16: ...B40 B20 Patient Monitor xii ...
Страница 17: ...1 Introduction ...
Страница 18: ......
Страница 27: ...2 System description ...
Страница 28: ......
Страница 58: ...B40 B20 Patient Monitor 2 30 ...
Страница 59: ...3 Installation ...
Страница 60: ......
Страница 69: ...4 Monitoring basic ...
Страница 70: ......
Страница 79: ...5 Alarms ...
Страница 80: ......
Страница 102: ...B40 B20 Patient Monitor 5 22 ...
Страница 103: ...6 Monitor setup ...
Страница 104: ......
Страница 123: ...7 Trends ...
Страница 124: ......
Страница 140: ...B40 B20 Patient Monitor 7 16 ...
Страница 141: ...8 Print and record ...
Страница 142: ......
Страница 153: ...9 Cleaning and care ...
Страница 154: ......
Страница 163: ...10 Troubleshooting ...
Страница 164: ......
Страница 181: ...11 ECG ...
Страница 182: ......
Страница 205: ...12 Impedance respiration ...
Страница 206: ......
Страница 215: ...13 Pulse oximetry ...
Страница 216: ......
Страница 232: ...B40 B20 Patient Monitor 13 16 ...
Страница 233: ...14 Non invasive blood pressure ...
Страница 234: ......
Страница 248: ...B40 B20 Patient Monitor 14 14 ...
Страница 249: ...15 Invasive blood pressure ...
Страница 250: ......
Страница 261: ...16 Temperature ...
Страница 262: ......
Страница 267: ...17 Airway gas ...
Страница 268: ......
Страница 293: ...18 Entropy ...
Страница 294: ......
Страница 316: ...Abbreviations A 12 ...
Страница 318: ...B40 B20 Patient Monitor B 2 ...
Страница 319: ......