9 Warnings and safety precautions
Warning!
!
•
In case of patients with implanted electronic device please carry out electrical stimulation
treatment only after having checked whether there is any risk.
•
Pieces of jewellery and glasses have to be taken off during the treatment.
•
Turn off cellular phones and radiophones or place them in a distance of 3 m from the
device.
•
Cardiac pacemakers can extremely be disturbed. In these cases the therapy should be
carried out only under continuous pulse and ECG control.
•
If the patient and/or the patient cable is in direct range of high frequency, shortwave or
microwave therapeutic device, a damage to the device or an injury to the patient cannot be
excluded. Please keep a distance of at least 3 m.
•
A simultaneous connection of the patient to a high-frequency surgery device can lead to
burns under the electrical stimulus electrodes.
•
The device is not designed for use in non-explosion-proof areas. If it is used in dangerous
anesthesia areas, an explosion cannot be excluded.
•
It is recommended not to exceed the current density of 2 mA/cm
2
on all electrode surfaces.
•
The electrodes must
adhere completely to the skin
in order to avoid
excessive local
current density
. Otherwise there can occur erythemas, and sometimes even
burns
.
•
In case of any visible failures please contact immediately
gbo
AG
or one of the service
agencies authorized by
gbo
AG.
10 Explanation of the signs used
Caution!
Observe the instructions for use !
Application part ungrounded, protection degree Type BF