Edit program
To move cursor, please turn
the knob below the display.
P9 S1
Lower frequency
0.1 Hz
Upper frequency
200 Hz
rise time
0s
signal time
0s
pause time
0s
Dynamics
off
Time
5 min
Figure 16: Edit program step menu
7. Using
the
modification knob
, select the parameter you want to change. Press
OK
to select this
parameter.
8. Modify the parameters as desired. The
OK key
stores your changes,
ESC key
leaves the parameter
unchanged.
9. When done, press the
ESC key
to return to the Edit program view.
!!
Note!
•
Surging is only reasonable at therapy frequencies greater than 2.25 Hz. If you try to modify the
surge parameters with frequencies below 2.25 Hz, a warning message will be displayed:
“Please note: Surge only with frequencies larger than 2.25 Hz!”
This message will also appear, if you try to decrease the frequency parameters below 2.5 Hz when
surge is activated.
•
Selecting Endosun 0° or Endosun 90° is done by choosing one of the frequencies parameter and
turning the
modification knob
to the right until “Endosun 0°” or “Endosun 90°” appears.
To delete single program steps, press the
softkey
[delete] when a program step is highlighted. Only the last
step will be erased. To delete the whole program, press the
softkey
[delete] when a program number is
highlighted. All steps will be erased then.
The loop parameter can be changed at the program step selection mode. Move the highlight to Loops,
confirm with the
OK key
, and modify its value straight forward.