![gbo DUODYNATOR 1049 Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/gbo/duodynator-1049/duodynator-1049_user-manual_3843093016.webp)
16
DUODYNATOR
®
1049
gbo
Medizintechnik
AG 2008
Version 1.3
3.3 Setting output intensity
3.3.1 Intensity regulators
The
Intensity regulators
serve for setting the output voltage. They are furnished in the form of a rotor pulse
generator. The current is increased by turning it right (in
+
direction) and it is reduced by turning it left
(in
-
direction). The referring numerical value is shown on the display.
Always turn the
Intensity regulator
slowly, until the current is clearly perceptible, but do not generate any
muscular contracture.
There is one regulator for each side of the device, a left one for channels I and II, and a right one for
channels III and IV. When in MFmod mode, there must be an activated channel when turning the
intensity
regulator
, otherwise the following message is displayed:
Select
channel!
In ICT mode, the respective
intensity regulator
is always activated as you do not need to select a channel.
When using surged programs, please note that increasing the intensity during the signal pause time will
cause the intensity to be automatically increased within a half second to the level of the signal time. Using
this method, changing intensity is always perceptible to the patient and can be changed at any time of the
surge cycle.
!!
Note!
•
After expiration of treatment time, the current to the patient is automatically switched off and the
intensity regulator reverts to
„0.0 V“
.
•
Before changing over from one current program to another, the intensity has to be set to
„0.0 V“
. Control the current indication on the display.
•
The menu can only be called up with the
Menu key
, if the intensity of all channels has been set to
„0.0 V“
.
•
When dynamic was active when the intensity was changed, dynamic will be turned off for 5
seconds and reactivated then.
3.3.2 Balance regulator
In ICT mode, the additional balance regulator will be displayed. It will be activated by pressing the respectiv
softkey
. Turn the
modification knob
slowly to shift the location of stimulation.
3.4 Changing treatment parameters
3.4.1 Treatment time
Depending on the type of current program the therapy time defined in the basic settings will be suggested.
To change the therapy time proceed as follows:
1. Press
Treatment time key
.