In accordance with the EMC-regulations for medical products we are obliged by law to 
provide the following information. 

Guidance and manufacturer’s declaration — electromagnetic emissions 

The equipment is intended for use in the electromagnetic environment specified below. The customer or 
the user of the equipment should assure that it is used in such an environment. 

Emissions test 

Compliance 

Electromagnetic environment – guidance 

RF emissions, 
CISPR 11 

Group 1 

The equipment uses RF energy only for its internal function. 
Therefore, its RF emissions are very low and are not likely to cause 
any interference in nearby electronic equipment. 

RF emissions, 
CISPR 11 

Class B 

Harmonic emissions, 
IEC 61000-3-2 (*) 

Class A 

Voltage fluctuation/flicker 
emissions, IEC 61000-3-3 (*) 

Complies 

The equipment is suitable for use in all establishments, including 
domestic establishments and those directly connected to the public 
low-voltage power supply network that supplies buildings used for 
domestic purposes. 

(*) Note: For devices with a power consumption between 75 W and 1000 W only. 

Guidance and manufacturer’s declaration — electromagnetic immunity 

The equipment is intended for use in the electromagnetic environment specified below. The customer or 
the user of the equipment should assure that it is used in such an environment. 

Immunity test 

IEC 60601- test level

Compliance level  Electromagnetic environment – 

guidance 

Electrostatic discharge (ESD), 
IEC61000-4-2 

±

6 kV  contact 

 

±

8 kV  air 

±

6 kV  contact 

 

±

8 kV  air 

Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material, the relative humidity 
should be at least 30 %. 

Electrical fast transient/burst,  
IEC 61000-4-4 

±

2 kV for power supply 

lines  
 

±

1 kV for input/output 

lines 

±

2 kV for power supply 

lines  
 

±

1 kV for input/output 

lines 

Mains power quality should be that of a 
typical commercial or hospital 
environment. 

Surge,  
IEC 61000-4-5 

±

1 kV differential mode 

 

±

2 kV common mode 

±

1 kV differential mode 

 

±

2 kV common mode 

Mains power quality should be that of a 
typical commercial or hospital 
environment. 

Voltage dips, short 
interruptions and voltage 
variations on power supply 
input lines,  
IEC 61000-4-11 
 

<5% U

τ

 

for ½ cycle 
(>95% dip) 
 
40% U

τ

 

for 5 cycles 
60% dip) 
 
70% U

τ

 

for 25 cycles 
30% dip) 
 
<95% U

τ

 

for 5 s 
(>5% dip) 
 

<5% U

τ

 

for ½ cycle 
(>95% dip) 
 
40% U

τ

 

for 5 cycles 
60% dip) 
 
70% U

τ

 

for 25 cycles 
30% dip) 
 
<95% U

τ

 

for 5 s 
(>5% dip) 
 

Mains power quality should be that of a 
typical commercial or hospital 
environment. 
 
If the user of the equipment requires 
continued operation during power mains 
interruptions, it is recommended that the 
equipment be powered from an 
uninterruptible power supply or a 
battery. 

Power frequency (50/60 Hz) 
magnetic field,  
IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic fields should 
be at levels characteristic of a typical 
location in a typical commercial or 
hospital environment. 

Note: U

τ

 is the a.c. mains voltage prior to application of the test level. 

Содержание DUODYNATOR 1049

Страница 1: ...gbo Medizintechnik AG 2008 Version 1 3 Interference Current Stimulation Device DUODYNATOR 1049 User Manual...

Страница 2: ...for errors or omissions All rights reserved No part of this manual may be reproduced in any form or by any means electronic mechanical or otherwise without the prior written permission of the gbo AG g...

Страница 3: ...anations Arial in type size 10 functions and control keys of the device Lucida in type size 10 or 11 text appears on the display of the device Warnings and safety precautions Warning Warnings which ha...

Страница 4: ...currents 9 2 START OF OPERATION 11 2 1 Transport and Assembly 11 2 2 Connection and switch on 11 2 2 1 Voltage selection and fuses 11 2 3 Basic settings 12 3 DESCRIPTION OF FUNCTION 13 3 1 Principle...

Страница 5: ...ble Check 26 6 ELECTRODES 27 6 1 Electrode Positioning 27 6 2 Single Pole Electrodes 28 6 3 Bipolar electrodes 28 7 BEHAVIOR IN CASE OF FAILURES 29 7 1 Failures 29 7 2 Application errors 29 8 MAINTENA...

Страница 6: ...ws the desired highest possible low frequency stimulation from the immediate area surrounding the electrode to be shifted into deeper regions since this is where it is generated Apart from the two dim...

Страница 7: ...o the keys LED key position flashing key was selected and remains active constantly lit key is selectable not lit key is not active and not selectable Table 1 Optical user support Furthermore an optic...

Страница 8: ...keys Int Int Mod DUODYNATOR 1049 I II III IV ESC Intensity regulator channels I II Treatment time key ESC key Modification knob OK key MENU key Intensity regulator channels III IV Figure 1 Device vie...

Страница 9: ...fting with the Modification knob 14 At the end of the therapy the triad gong will sound The current to the patient will be automatically reduced 1 6 Short instructions treatment with MFmod programs si...

Страница 10: ...10 DUODYNATOR 1049 gbo Medizintechnik AG 2008 Version 1 3 13 At the end of the therapy the triad gong will sound The current to the patient will be automatically reduced...

Страница 11: ...onnection and switch on The basic setting of the current stimulation device has been set for a supply voltage of 230 V If needed the device can be switched over to the following supply voltages 230 V...

Страница 12: ...ss the menu key Turn the modification regulator to move the cursor to Settings If the default language is active move the selection to Einstellungen Now press the OK key to step down on menu level Mov...

Страница 13: ...and dynamics parameter 3 1 4 Current programs The user can easily change the current program of each channel Current programs are divided into Predefined programs such as Lipolysis The name of the pro...

Страница 14: ...tration of drugs Here the procedure is the same as in case of galvanization The quantity of the penetrated drug depends on the purity of its solution the size of the electrodes the intensity of curren...

Страница 15: ...program see chapter 3 1 4 Current programs for details press the Softkey Menu Using the modification knob you may select the required program Press the OK key to load a program ESC key to abort select...

Страница 16: ...pause time will cause the intensity to be automatically increased within a half second to the level of the signal time Using this method changing intensity is always perceptible to the patient and can...

Страница 17: ...oceed as follows 1 Activate the channel by pressing the respectiv channel key also when in ICT mode 2 By pressing the OK key you may step through the parameters The currently selected parameter is dra...

Страница 18: ...time key 3 5 3 OK key The OK key has several functions when in menu it is used to step down to the next menu level or to select a parameter to be changed in therapy mode it is used to activate the sur...

Страница 19: ...u Figure 10 Softkey 3 5 7 Interference current With the ICT key on either side of the current stimulation device can be switched over into ICT mode This can only be done when the intensity on both cha...

Страница 20: ...are shorted out by a conductive foreign substance This could be a metallic spectacle frame in case of tinnitus treatment If during the treatment that shunt is removed e g by taking off the glasses thi...

Страница 21: ...ram Edit program Settings Cable Check To move cursor please turn the knob below the display Figure 13 Menu Note The menu can be chosen with the Menu key only if the intensity is 0 0 V in all channels...

Страница 22: ...lue 8 Confirm by pressing the OK key If you do not want to change the parameter press the ESC key 9 You can change other values by selecting the parameter with the OK key as you did before 10 You can...

Страница 23: ...200 Hz 5 min S4 S5 Figure 14 Load program menu 5 Turn the modification knob to choose a specific program You can choose out of 30 programs You will see the program details on the right display half 6...

Страница 24: ...ey 2 Move the scroll bar with the modification knob to Edit program 3 Confirm by pressing the OK key 4 The contents of the display is similar to the Load program menu Use the modification knob to sele...

Страница 25: ...to modify the surge parameters with frequencies below 2 25 Hz a warning message will be displayed Please note Surge only with frequencies larger than 2 25 Hz This message will also appear if you try t...

Страница 26: ...ue 6 To select a channel please turn the Modification knob 7 Press the OK key to start the test 8 Move and twist the cable at all parts in order to detect a weak connection or disruption 9 If the cabl...

Страница 27: ...cket so that the pocket protrudes on the open side with about 1 cm 3 After multiple use the electrode intermediate layer or electrode pocket can slightly shrink If the distance of about 1 cm is no lon...

Страница 28: ...and standard Electrode pockets are necessary for application and fixing material for bandage see accessories Each bipolar electrode set is placed into a well wetted pocket so that the black conductiv...

Страница 29: ...7 1 Failures Checking power supply ERROR Suggestions Switch off and on again If the failure occurs again the device is out of function Please contact a service agent authorized by the manufacturer Che...

Страница 30: ...of at least the following criteria Electrical safety check in accordance with the test plan of the manufacturer Check of the device in respect of external integrity Check of all display and operating...

Страница 31: ...damage to the device or an injury to the patient cannot be excluded Please keep a distance of at least 3 m A simultaneous connection of the patient to a high frequency surgery device can lead to burn...

Страница 32: ...ter IPX1 Dimensions max 15 cm x 41 cm x 48 cm Height x Depth x Width Weight max 11 kg without accessories Color RAL 9002 Display LCD backlighted 240 x 128 dots full graphic Environmental conditions Op...

Страница 33: ...nductive rubber electrode 80 mm x 120 mm pack of 2 017 0 0047 Conductive rubber electrode 115 mm x 175 mm pack of 2 017 0 0046 Self adhesive electrode 50 mm x 57 mm pack of 4 017 0 0021 Self adhesive...

Страница 34: ...ew 15 Dynamic 14 E Electrodes 28 elektrode fleece 31 elektrode positioning 28 F function keys 8 G Galvanic current 15 graphic display 8 I Intensity 17 Intensity monitoring 21 iontophoresis 15 L LED 8...

Страница 35: ...30 min Edema 10 Hz Pain Endosun 0 10 seconds Endosun 90 10 seconds 90 110 Hz remaining time Steps 1 2 combined for MFmod therapies Universal Endosun 0 10 seconds Endosun 90 10 seconds 0 1 200 Hz Step...

Страница 36: ...rs DUODYNATOR 1049 is an electronic device For its disposal the according regulations for electronic devices have to be observed Incidentals have to be disposed with residual waste On request the manu...

Страница 37: ...munity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concre...

Страница 38: ...to 800 MHz d 2 3 P for 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts according to the transmitter manufacturer and d is the recommended separation distance...

Отзывы: