5.4 Cable Check
If you are unsure about the functionality of a patient cable (for instance loose connections, which may occur
on heavy use), you may verify the correct function using the cable check.
1. Press the
MENU key
.
2. Move the scroll bar with the
Modification knob
to Cable Check.
3. Confirm by pressing the
OK key
.
4. Proceed as shown in the display:
Please check that nobody is connected to the device! Connect cable (patient cable) to be tested into a
black and a colored plug of the distribution panel!
5. Press
the
OK key
to continue.
6. To select a channel, please turn the
Modification knob
.
7. Press
the
OK key
to start the test.
8. Move and twist the cable at all parts in order to detect a weak connection or disruption.
9. If the cable and all connections are okay, the message OK is displayed after 20 seconds.
10. Otherwise ERROR is displayed. The patient cable is defective, please replace it.
11. The distribution cable can also be tested:
•
Please use a patient cable, proved to be OK.
•
Use this cable for testing other distribution cables. If the message ERROR is displayed, the
distribution cable is damaged and must be replaced.
12. Exit the Menu Cable Check
by pressing the
ESC key
.
13. On the display the Main menu appears again.
14. Exit the Main menu by pressing the
ESC key
.
15. On the display the therapy display appears again.