![Gatan UltraScan 894 Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/gatan/ultrascan-894/ultrascan-894_manual_569111009.webp)
UltraScan Use
r
’s Guide Reference Series Rev 1.0
2-1
2 Getting Started
The following steps will enable the first-time user to switch on the digital
camera, launch DigitalMicrograph and prepare for the acquisit ion of an image.
•
Make sure that the viewing chamber is at the proper vacuum.
•
Make sure the water and the three color-coded pressure lines are
correctly connected to the camera and proper water pressure is
maintained.
•
Switch on the CCD camera controller (Fig. 2-1), the LED will turn
green.
•
Flip the Shutter switch to AUTO.
•
Flip the Peltier Cooler switch to the COOL position. For best results it
is recommended waiting (2 hours or more) for the system to stabilize.
•
Turn the computer on (if it is not on yet) and double click on the
DigitalMicrograph icon to launch the application.