UltraScan Use
r
’s Guide Reference Series Rev 1.0
4-1
4 Quick Reference
The steps necessary for operating Gatan CCD camera are summarized here.
•
Make sure that the viewing chamber is at the proper vacuum.
•
Make sure the water and the three color-coded pressure lines are
correctly connected to the camera and proper water pressure is
maintained.
•
Switch on the CCD camera controller.
•
Flip the Peltier Cooler switch to the COOL position. For best results it
is recommended waiting (2 hours or more) for the system to stabilize.
•
Flip the shutter switch to AUTO.
•
Turn the computer on (if it is not on yet) and double click in the
DigitalMicrograph icon to launch the application.
•
Check the TEM session information by clicking on the
Global Info
button on the
Image Info
palette (Figure 3-3).
•
Run the Set Bias Levels routine by selecting the
Set Bias Levels
button
under the
Camera
menu.
•
Prepare a quadrant correction reference by choosing
Prepare
Quadrant Correction Reference
under the
Camera
menu (Figure 2-
3).
•
Prepare a gain reference image by choosing
Preparing Gain
Reference
under the
Camera
menu (Figure 2-3).
Содержание UltraScan 894
Страница 21: ...Magnification Calibration UltraScan User s Guide Reference Series Rev 1 0 2 13 Figure 2 15 Calibration window...
Страница 30: ...Image Saving and Printing UltraScan User s Guide Reference Series Rev 1 0 3 9 Figure 3 10 Control palette...
Страница 46: ...A2 Description UltraScan User s Guide Reference Series Rev 1 0 VIII Figure A 7 Controller s back panel...
Страница 55: ...UltraScan User s Guide Reference Series Rev 1 0 XVII Appendix 2 CE certification...