Set Bias Levels
UltraScan Use
r
’s Guide Reference Series Rev 1.0
2-4
Figure 2-3
(a) Camera Menu
(b) Camera configuration
•
Select
Configure Camera
in the
Camera
menu (Figure (2-3a).
•
In the Camera configuration dialog box (Figure 2-3b), you can change
the name of the camera by typing in the
Name
field.
•
Image rotation can be specified in the
Rotation
pull-down menu.
•
The coordinates for column defect correction are entered in the
Defects
field. For example, entering the following:
r0-8, r4088-4095, c0-8,c4088-4095
(r and c indicates row and columns respectively) corrects for the 8
pixels wide edge which does not contain active CCD pixels in the case
of the UltraScan 4000.
•
It is important to specify the
Primary
and
Alternate
(if there is one set
up) shutter available to the camera.
•
Finally you should specify the shutter status, i.e. the state of the shutter
when the CCD camera is not acquiring any images.
Shutter Normally
Closed
is the default setting.
2.3 Set Bias Levels
In 4-channel CCD readout, the entire CCD sensor is divided into 4 sectors
(called quadrants). It is common that each quadrant has a different bias level.
Содержание UltraScan 894
Страница 21: ...Magnification Calibration UltraScan User s Guide Reference Series Rev 1 0 2 13 Figure 2 15 Calibration window...
Страница 30: ...Image Saving and Printing UltraScan User s Guide Reference Series Rev 1 0 3 9 Figure 3 10 Control palette...
Страница 46: ...A2 Description UltraScan User s Guide Reference Series Rev 1 0 VIII Figure A 7 Controller s back panel...
Страница 55: ...UltraScan User s Guide Reference Series Rev 1 0 XVII Appendix 2 CE certification...