5 Care and Maintenance
This chapter contains some general tips on the care and maintenance of your
UltraScan camera. It also contains a list of questions that are often asked by
users.
5.1 General Precautions
The following are some general precautions users should adhere to in the care
and maintenance of their camera.
5.1.1 Peltier cooler maintenance
•
In standard operation conditions the temperature display is set to a
temperature of -25ºC. The temperature should never exceed 50ºC. A
LED above the reset switch is illuminated when an over-temperature
condition is detected.
The first light controller monitors the temperature of the Peltier heat sink. If the
temperature exceeds 50ºC an automatic protection will shut down the cooler.
Cooling can be restarted by pressing the reset button on the First Light
controller or by powering the controller off and on. The temperature of the
Peltier heat sink can be checked by pressing the “Heatsink” switch.
•
The Peltier cooler should not be switched on if the cooling water is not
flowing.
The Peltier cooler consumes about 20 W of power. If this heat is not removed
by the cooling water, an over-temperature condition will be detected and the
automatic protection will shut down the cooler.
•
Check the flow of cooling water periodically.
Содержание UltraScan 894
Страница 21: ...Magnification Calibration UltraScan User s Guide Reference Series Rev 1 0 2 13 Figure 2 15 Calibration window...
Страница 30: ...Image Saving and Printing UltraScan User s Guide Reference Series Rev 1 0 3 9 Figure 3 10 Control palette...
Страница 46: ...A2 Description UltraScan User s Guide Reference Series Rev 1 0 VIII Figure A 7 Controller s back panel...
Страница 55: ...UltraScan User s Guide Reference Series Rev 1 0 XVII Appendix 2 CE certification...