Solarus 950 Owner’s Manual and User’s Guide
43
5.
Slide the pump out of the cabinet until the two connectors on the top can be
accessed. Unplug the two connectors (one is held captive by a screw).
6.
Slide the pump completely out of the cabinet.
7.
The pump can be installed by following the reverse of these directions.
Removing the Molecular Drag Pump
1.
Shut off the power to the system and unplug the power cord. The power
switch is located on the rear of the cabinet near the right side.
2.
Remove the front bezel and both covers. (see instructions above)
3.
Disconnect the backing line from the pump by loosening the knurled knob
and pulling (not the hex nut).
4.
Unscrew the rubber foot and remove the thermistor.
5.
Unscrew the bolts on the three KF clamps that hold the pump to the manifold.
Hold the pump as this is done to ensure that it does not fall.