44
⑥
Perform leak check. If there is no air leakage, the
replacement is completed. For details, refer to
“7.6 Leak Check”.
8.2 Checking the Total Operating Time of the Air Pump
The total operating time of the built-in air pump is recorded and can be checked by the user. Use this
information as a guide when using more than one GSP-501FT properly or to know when the
maintenance time.
①
While pressing the “MODE Button”, press the
“POWER Button”.
Press
while pressing
②
“Pt” (Pump Total Operation Time) is displayed in
hour.
(Example) Pump total operating time:
10 hours
③
After checking, press and hold the “POWER
Button” to exit. (The power is turned off.)
Press and Hold
△
Note
The total operating time of the built-in air pump is displayed by rounding down minutes or
less.
(Example: When the pump operation time is 4 hours 58 minutes, indicated value is 4 hours.)
Refer to “11.3 Recommended Time for Maintenance Parts Replacement” for the
recommended replacement timing of the air pump.