19
When the instantaneous flow rate is set to 500 mL/min
Min: 0.01L Max: 900 L (=500mL/min×30 h)
①
When the pump is turned on by pressing the
"POWER Button", the flow rate conversion
temperature is displayed for 2 seconds, and then the
sampling results in the previous operation are
displayed.
Previous sampling result
②
Press "MODE Button" to select the Volume Mode.
Press
twice
↓
③
Check that the set value of "Instantaneous flow rate"
is blinking, and set the desired instantaneous flow
rate with the "INCREASE Button" and
"DECREASE Button". Press and hold to make fast
forward.
[Setting range]
10
~
500mL/min
※The setting
increment is 1mL/min.
※When the instantaneous flow rate is set at 10 to
49mL/min, the pump runs intermittently with the
flow rate of 50mL/min. Refer to "6.12 Intermittent
Operation" for details.
(Example) Flow rate: 100 mL/min
↓
Press
or