To set the sampling time, press the "SELECT
Button", to make the "Time" setting value blink,
and set the desired sampling time with the
"INCREASE Button" and "DECREASE
Button”. Press and hold to make fast forward.
[setting range] 0:01 to 30:00 (1 min to 30 h)
※The setting
increment is one-minute.
※When the instantaneous flow rate and
sampling time are set, the integrated volume is
automatically calculated, and the value is
displayed at the "Integrated volume" position.
※Set to "
" for continuous sampling.
Setting procedure
When 0:01 is displayed, press "DECREASE
Button” once.
When 30:00 is displayed, press
"INCREASE Button” once.
(When "Time" is set to "
", "
" is also displayed at the
"Integrated volume” position)
Press
↓
(Example) Sampling time: 10 min.
↓
Press
or
⑤
To set the standby time, press "SELECT Button"
to make the "Standby" setting value blink, then
set the desired standby time with the
"INCREASE Button" and "DECREASE
Button". Press and hold to make fast forward.
[setting range] 0:00 to 24:00 (0 min to 24 h)
※The setting
increment is one-minute.
Press
↓
(Example) Standby time: 5 min.
↓
Press
or
⑥
Check that the instantaneous flow rate, sampling
time, integrated volume, and standby time are set
correctly.