The screen is switched by pressing the "SELECT
Button" and the settings (mode, instantaneous
flow rate, sampling time, integrated volume, and
standby time) for the completed sampling
operation are displayed.
Press
↓
⑩
Press and hold the "POWER Button" to turn off
the power.
Press and Hold
△
Note
To repeat sampling with the same settings, press and hold the "Start/Stop Button" in the state
described in
⑧
and
⑨
above to start the sampling operation.
6.4 Volume Mode
△
Note
The integrated volume rate cannot be set to beyond 30 hours of sampling.
See the settable range in the graph below.
Example)
When the instantaneous flow rate is set to 10 mL/min (intermittent run)
Min: 0.01L Max: 18 L (=10 mL/min x 30 h)