33
③
Check that "tStd" is blinking.
④
Press the "INCREASE Button" and
"DECREASE Button" to select the desired
temperature between 20°C and 25°C.
Press
or
⑤
Press "START/STOP Button" again. If the
displayed temperature blinks three times, the
setting is completed.
Press
↓
⑥
To make other settings continuously, press
"MODE Button" to select the desired item.
If not, press and hold the "POWER Button" to
exit the calibration mode. (The power is turned
off.)
[For other settings]
Press
[To exit the calibration mode]
Press and Hold
△
Note
If no operation is performed for 2 minutes while selecting the flow rate conversion
temperature (in
③
and
④
above), "
" is displayed and the display returns to the state
of the item selection in the calibration mode without saving any changes.
Press the "MODE Button" while selecting the flow rate conversion temperature (in
③
and
④
above) to exit the "Setting of Flow Rate Conversion Temperature" without saving any
changes. In this case, "
" is displayed.