36
7.5 Zero Point Adjustment
Adjust the zero point of GSP-501FT built-in flowmeter.
①
(If it is not in the calibration mode), activate the
calibration mode and make sure that "25°C tStd"
is displayed. (When the conversion temperature is
set to 20°C, "20°C tStd" is displayed.)
②
Press "MODE Button" twice and confirm that
"0AdJ" is displayed.
Press twice
↓
③
Press "START/STOP Button".
Press
④
The display of blinking "0AdJ"
→
Zero point of
flow rate prior to calibration
→
blinking "0AdJ"
→
blinking "0mL/min" (Zero point of flow rate
after calibration) indicates the completion of zero-
point adjustment. If the zero point is 0mL/min
after calibration, it is normal. If it does not reach 0
mL/min, recalibration is necessary.