CMM USER'S MANUAL (E2.51)
48
Message
Explanation
CMM system log created.
Creation date of the file and first entry.
Starting CMM Version: <Version_Number>
CMM version information
Error - Cannot communicate with test gas
generator: <code> <description>
There was an error communicating with the test gas generator,
check IP, power and Ethernet cable.
Error - Cannot communicate with analyzer:
<code> <description>
There was an error communicating with the analyzer, check IP,
power and Ethernet cable.
Hg
0
Span calibration error, result: X, target:
Y±Z.
Hg
0
span calibration has failed. System servicing is possibly
required.
Hg
0
Span check error, result: X, target: Y±Z.
Hg
0
span check has failed. System servicing is possibly required.
HgCl
2
Span check error, result: X, target:
Y±Z.
HgCl
2
span check has failed. System servicing is possibly required.
Hg
0
Linearity check error, result: X, limit: ±Y.
Linearity check has failed. System servicing is possibly required.
Zero calibration error, result: X, limit: ±Y.
Zero calibration has failed. System servicing is possibly required.
User mode changed to <user mode>
User access level information
Standby mode activated, measurement
stopped
Ongoing measurement will stop automatically if Standby mode is
activated. In case of continuous measurement, measurement will
resume automatically when standby mode is deactivated.
Instrument air pressure is too low,
measurement stopped
Measurement stops automatically if instrument air pressure is too
low. Check instrument air pressure level.
Measurement stopped, Cell Temp
Alarm=%d, Line Temp Alarm=%d, Probe
Temp Alarm=%d
Ongoing measurement will stop automatically if any of these alarms
is active (Line temperature, Cell temperature, Probe temperature
alarm). In case of continuous measurement, measurement will
resume automatically if these alarms are not active anymore.
1 = Active alarm, 0 = Alarm is not active
Resuming measurement
Resuming continuous measurement
Hg0 Bath temperature is not within limits.
value: %.2f, Low: %.2f, High: %.2f
Hg0 bath temperature is not within defined limits.
Photomultiplier communication related issues
Cannot close COM port handle!
There is an error handling the photomultiplier RS232
communications port, the device is not responding when cancelling
a measurement or exiting the program.
Invalid COM port handle!
There is an error handling the photomultiplier RS232
communications port, the device is not responding when starting a
measurement or reinitializing the PMT.
PMT Gate Status Error
There is an error handling the photomultiplier RS232
communications port, Restart the analyzer. If it doesn’t help,
contact the service.
Cannot create PMT stop event!
Cannot create PMT control thread!
Cannot set COM port timeouts!
Cannot set COM parameters!
Cannot turn off PMT high voltage!
Cannot turn on PMT high voltage!
Cannot kill PMT count disabling!
Cannot deactivete PMT gate!
Cannot set PMT number!
Cannot set PMT counting time!
Cannot start PMT!
Error in PMT read data!
Table 18. All possible log messages