CMM USER'S MANUAL (E2.51)
13
1.2.9.1 Heated probe tube
Material:
SS 316, glass coated sample wetted parts
Temperature:
Maximum setting 250 °C
Length:
122 cm
Mounting flange:
DP100PN16
Other materials, lengths and temperatures are available on request.
1.2.10 Heated line
Tube size:
2 * 6/8 mm
Core material:
PFA Teflon core
Temperature:
Maximum 200 °C
Fittings:
8 mm Swagelok
Power supply:
230 VAC or 115 VAC
Power density:
200 watts/meter
Dilution and blowback air:
Unheated 2 * 4/6 mm Teflon core, 6 mm Swagelok
The maximum length for the heated line is 35 m with 230 VAC and 14 m with 115 VAC power supply.
Analyzer and test gas generator are connected to dilution probe with combined heated line which
divides into two parts on both ends.
1.2.11 Electrical connections
Main supply:
3 * 16 A (2 – 14 m sample line)
3 * 32 A (15 – 30 m sample line)
Power consumption:
~ 8 kW (the full CMM with heated lines, 25 m)
1.2.12 Measuring data outputs
The CMM system is equipped with 4 analog outputs representing the result total Hg concentration with
different ranges.
Analog output range:
4 – 20 mA. Active, load 350
Ω
max.
1.2.13 Enclosure
Material:
Bake painted steel
Dimensions (cm):
212 * 61 * 70 (cooling unit on top)
Weight:
230 kg (the full CMM cabinet)
27 kg (dilution probe + probe tube)
Protection:
IP54