CMM USER'S MANUAL (E2.51)
17
Module
Pin
Signal
Information
B10 KL9186
1
Stand by
Control by free contacts. Closed = standby
B11 KL1408
1
Stand by
Control by free contacts. Closed = standby
B14 KL2612
1
System alarm
Free contacts. Alarm = open
B14 KL2612
3
System alarm
Free contacts. Alarm = open
B14 KL2612
2
Service request
Free contacts. Alarm = open
B14 KL2612
4
Service request
Free contacts. Alarm = open
B15 KL2612
1
Maintenance
Free contacts. Maintenance on = open
B15 KL2612
3
Maintenance
Free contacts. Maintenance on = open
B15 KL2612
2
Concentration alarm
Free contacts. Alarm = open
B15 KL2612
4
Concentration alarm
Free contacts. Alarm = open
B16 KL2612
1
Results valid
Free contacts. Results valid = closed
B16 KL2612
3
Results valid
Free contacts. Results valid = closed
B16 KL2612
2
O
2
Sensor Calibration
Free contacts. Calibrate = closed
B16 KL2612
4
O
2
Sensor Calibration
Free contacts. Calibrate = closed
B12 KL4414
1
Hg concentration, analog o
4 – 20 mA, load max 350
Ω
B12 KL4414
3
Hg concentration, analog output1-
4 – 20 mA, load max 350
Ω
B12 KL4414
2
Hg concentration, analog o
4 – 20 mA, load max 350
Ω
B12 KL4414
4
Hg concentration, analog output2-
4 – 20 mA, load max 350
Ω
B12 KL4414
5
Hg concentration, analog o
4 – 20 mA, load max 350
Ω
B12 KL4414
7
Hg concentration, analog output3-
4 – 20 mA, load max 350
Ω
B12 KL4414
6
Hg concentration, analog o
4 – 20 mA, load max 350
Ω
B12 KL4414
8
Hg concentration, analog output4-
4 – 20 mA, load max 350
Ω
B13 KL3054
1
O
2
concentration
4 – 20 mA
B13 KL3054
3
O
2
concentration
4 – 20 mA
X2
L1
Mains in L1
230/400 VAC, 16 A, TN-S, 2.5-6.0 mm
2
X2
L2
Mains in L2
230/400 VAC, 16 A, TN-S, 2.5-6.0 mm
2
X2
L3
Mains in L3
230/400 VAC, 6-32 A, TN-S, 2.5-6.0 mm
2
X2
N
Mains in N
230/400 VAC, 6-32 A, TN-S, 2.5-6.0 mm
2
X2
PE
Mains in PE
230/400 VAC, 6-32 A, TN-S, 2.5-6.0 mm
2
X2
1
TcK+
Heated line
X2
2
TcK-
Heated line
X2
3
PE
Heated line
X2
4
L
Heated line
X2
5
N
Heated line
XN1
Ejector motive gas to probe
6 mm OD tube, Swagelok
XN2
Blowback air to probe
6 mm OD tube, Swagelok
XN3
Instrument air in
6– 8 bars, 8 mm OD tube, Swagelok, 60 l/min
XE
PE
Potential equalizer
Potential equalizer bar on the right hand side of the
cabinet
Table 1. Connector plate connections
2.5.1 Mains
Only professional electrician is allowed to do the connections!
Main supply for the CCM is 230/400 VAC 3*(16-32) A TN-S. The conductor area of an electrical cable
is 2.5-6 mm
2
and it is depended on the length of the electrical cable and sample line. Electricity and
protective earth are connected to connector bar X2. Do not connect power to the system before all
electrical and gas connections are properly made.