CMM USER'S MANUAL (E2.51)
40
It is needed to make the calibration until the system is warmed up. Calibration can be performed
manually but the system can be also set to make automatic calibration after first measurement.
After calibration, the measurement can be placed on the
Measure
button
.
4.4 Service and maintenance during the use
In order for the system to operate reliably, it is important that it is correctly maintained.
4.4.1 Maintenance plan
The maintenance required depends strongly on the application (Table
11
). The maintenance work
procedures are described in this manual.
Maintenance interval
Maintenance work
Regularly
System visual inspection
Approximately 6 months
Inspection of the probe filter plate
Approximately 6 months
Change of the Hg lamp
Approximately 3 months
Adding HgCl
2
liquid to container, if needed ((CMM AutoQAL))
Approximately 12 months
Replacement of the dryer materials
Approximately 12 months
Replacement of the cartridge filters
Approximately 12 months
Change of the vacuum pump, depending on condition
Approximately 5 years
Change of the probe filter plate
Table 11. Maintenance plan