CMM USER'S MANUAL (E2.51)
27
zero indicates blowback is not performed during continues
measurement.
Blowback time
Defines the time of day when first blowback should be
performed.
Pressure calibration group
Pressure calibration
x mbar
Calibration value for sample cell pressure sensor.
Temperature settings group
Lamp**
Defines the lamp temperature setpoint.
Oven**
700 °C
Defines the converter oven temperature setpoint.
Cell**
180 °C
Defines the sample cell temperature setpoint.
Line**
180 °C
Defines the heated line temperature setpoint.
Table 6. Analyzer settings (example parameters only, not for reference)** – Read-only in basic user
mode but editable in service mode
3.4.2 Settings of the test gas generator
The test gas generator settings control the operation of the test gas generator and the various
calibrations performed (
Figure 18
). The controls are explained in detail in the table below (Table
7
).
Figure 18. Test gas generator settings (example parameters only, not for reference)