CMM USER'S MANUAL (E2.51)
26
no alarms on its display. Switch on the test gas generator and the analyzer both at the same time
(switches are on the front panel).
After switching on the power, check that all settings are correct. The heating process of the system
takes about 2 hours.
3.4.1 Settings of the analyzer
The analyzer settings control the operation of the analyzer and the continuous measurement cycle
(Figure
17
). The controls are explained in detail in the table below (Table
6
).
Figure 17. Analyzer settings
Parameter
Value
Description
Measuring time group
Measuring time
x min
Defines the measuring time. Please note that the measuring
time must be a multiple of 1 minute.
Flush time
x min
Defines sample flushing time before first measuring on
continuous mode.
Blowback group
Perform before measure
Yes / No
If ticked, performs a blowback of the probe filter before starting
or resuming a measurement sequence.
Perform before calibration
Yes / No
If ticked, performs a blowback of the probe filter before starting
a calibration sequence.
Blowback duration
0 – X s
Duration of the blowback.
Interval
0 - X min
Defines the time interval of automatic blowbacks during normal
continuous measurement. The next blowback is performed
after the time interval has elapsed from the previous blowback.