Chapter 6: Adding a new Load/Power Supply to Framework
6-20
Key Name
Type
Required
Description
CalCellValue
Float
YES
This is the calibration cell value, in ohms.
CalCell
String
YES
This is the description of the calibration cell which is
used to tell the user what connection to make.
VchRange
Float
YES
This is the voltage, in volts, used to set the Vch range on
the IFC1010 instrument. For this step, the range is
usually set to the to a gain which matches the voltage
expected at the cell. A x1 gain or the 3 V range will give
full scale for the LPI. A x10 gain or the 300 mV range will
give 1/10 full scale. A x100 gain or the 30 mV range will
give 1/100 full scale.
IchRange
Float
YES
This is the voltage, in volts, used to set the Ich range on
the IFC1010 instrument. For the calibration, the range is
usually set to the x1 gain or the 3 V range.
IAppliedDCI
Float
YES
This is the current, in Amps, applied to the cell.
TApplied
Float
YES
This is the time, in seconds, that
IAppliedDCI
is
applied to the cell.
TRest
Float
YES
This is the time, in seconds, that the calibration routine
waits before moving to the next step. This is to help
alleviate any temperature buildup in the calibration
resistor.
Enabled
Integer
YES
The enabled field is used to tell the calibration routine if
this calibration is active for the power device. If the
calibration is disabled, the previous values are not used.
0 = Disabled
1 = Enabled
Содержание LPI1010
Страница 2: ...ii...
Страница 6: ......
Страница 14: ......
Страница 30: ...Chapter 4 Hardware Installation 4 10...
Страница 36: ......
Страница 62: ......
Страница 64: ......
Страница 66: ......
Страница 67: ...Appendix D Certificate of conformity Appendix D VII Certificate of conformity...
Страница 68: ...Appendix D Certificate of conformity Appendix D VIII...