Chapter 5: Calibration
5-5
Figure 5-8
Entering Resistor value of the Calibration Cell
9)
The following steps require to turn the output switch of your electronic load or programmable power
supply ON and OFF several times. Follow these instructions to continue the calibration procedure.
10)
During the LPI calibration, you are required to switch the calibration cell. A new window appears with
installation instructions and the required calibration cell, see. In this case, a 0.2 Ω cell is required. Always
verify that the output switch of the electronic power supply or programmable power supply is turned
OFF
before installing the calibration cell. Press
OK
after installing the cell.
Figure 5-9
Entering Resistor value of the Calibration Cell
11)
Again, the calibration procedure asks you several times to turn the output switch of your electronic load
or programmable power supply ON and OFF. Follow these instructions to continue the calibration
procedure.
12)
After the calibration is done, a window appears to confirm its completion. Click the
OK
button to
acknowledge completion.
Figure 5-10
Completed Calibration
The LPI1010 D-sub Module contains an EEPROM memory chip. All cell cables with EEPROM memory are
known as
dynamic cables
. After performing a calibration, the calibration results are stored in the cable.
You can also see all calibration results in the Calibration pane at the bottom of the Gamry Instrument
Manager, see Figure 4-9.
Содержание LPI1010
Страница 2: ...ii...
Страница 6: ......
Страница 14: ......
Страница 30: ...Chapter 4 Hardware Installation 4 10...
Страница 36: ......
Страница 62: ......
Страница 64: ......
Страница 66: ......
Страница 67: ...Appendix D Certificate of conformity Appendix D VII Certificate of conformity...
Страница 68: ...Appendix D Certificate of conformity Appendix D VIII...