Appendix C : I/O Connectors--User I/O Connector
Appendix C - 1
Appendix C : I/O Connectors
The Interface 1000 includes several I/O connectors other than the Cell Connector described in Appendix B.
The details of these connectors are described here.
User I/O Connector
This connector contains a number of signals used to interface the Interface 1000 to external apparatus. It is the
miniature 15-pin female D-shaped connector on the rear panel of the Interface 1000.
The pin out of this connector is shown in Table C-1.
Table C-1
User I/O Connector
Pin
Name
Use
1
Analog Output High
The auxiliary output signal (DAC output)
2
Analog Output Low
The auxiliary output ground connection (ground)
6
Ground
Digital ground
7
Digital Out 0
A CMOS-compatible digital output; connecting device input
current must be below 10 µA
8
Digital Out 1
A CMOS-compatible digital output; connecting device input
current must be below 10 µA
9
Digital Out 2
A CMOS-compatible digital output; connecting device input
current must be below 10 µA
10
Digital Out 3
A CMOS-compatible digital output; connecting device input
current must be below 10 µA
11
Digital In 0
A CMOS-compatible digital input; 2.2 kΩ input impedance
12
Digital In 1
A CMOS-compatible digital input; 2.2 kΩ input impedance
13
Digital In 2
A CMOS-compatible digital input; 2.2 kΩ input impedance
14
Digital In 3
A CMOS-compatible digital input; 2.2 kΩ input impedance
15
+5 V
Power; 50 mA maximum current
Caution:
Floating operation of the Interface 1000 can be compromised by improper
cables to the
User I/O
connector. We do not recommend use of standard 15-pin shielded cables with this
connector. Custom cables with the shield connected to pin 6 of the D-connector are preferred.
Caution:
The ground pin (pin 6) on this connector is earth ground and the D-connector
shell is the Interface 1000
Floating Ground
. Improper
User I/O
connections can damage an Interface 1000
that is connected to a high-energy cell containing earth-grounded terminals.
Содержание Interface 1000
Страница 8: ......
Страница 18: ......
Страница 36: ...Chapter 4 Cell Connections Membrane Cell Connections 4 4...
Страница 44: ...Chapter 5 Panel Indicators and Connectors Rear Panel 5 8...
Страница 62: ...Chapter 8 Measurement of Small Signals Floating Operation 8 8...
Страница 68: ...Appendix A Interface 1000 Specifications Floating Operation Appendix A 6...
Страница 70: ...Appendix B Interface 1000 Cell Connector Floating Operation Appendix B 2...
Страница 78: ...Appendix E CE Certificate Appendix E 2 Low Voltage Certificate of Conformance...
Страница 79: ...Appendix E CE Certificate Appendix E 3 RFI Certificate of Conformance...
Страница 80: ...Appendix E CE Certificate Appendix E 4...
Страница 84: ...Appendix F Heat in Interface 1000 MultEchem Systems Appendix F 4...