Chapter 6: Instrument Circuitry--Interface 1000 Schematic/Block Diagrams
6 - 4
Figure 6-3
One A/D Signal Chain in the Interface 1000
Notes for Figure 6-3:
This diagram shows one of two identical ADC channels. One channel is dedicated to measurement of
the potentiostat’s current signal and the other is used to measure the voltage signal.
Both A/D converters are triggered simultaneously to start a conversion. This trigger and the pulse
updating the Scan DAC voltage are under the control of a hardware state-machine. This insures that all
waveform and data-acquisition timing is tightly controlled and reproducible point-to-point.
By default, the data acquisition is synchronized with the system’s 600 kHz power-supply switching-
frequency reducing noise from the power supply. Data-acquisition times that are a multiple of 1.666
µs will maintain this synchronization.
All analog signals are received differentially as shown here.
The 5 Hz, 1 kHz, 10 kHz, 200 kHz and 1.4 MHz filters are two-pole Butterworth filters.
The voltage-measurement channel includes a programmable 1×−0.25× gain circuit. In the 0.25 gain
position, voltages as high as 12 V can be measured.
Содержание Interface 1000
Страница 8: ......
Страница 18: ......
Страница 36: ...Chapter 4 Cell Connections Membrane Cell Connections 4 4...
Страница 44: ...Chapter 5 Panel Indicators and Connectors Rear Panel 5 8...
Страница 62: ...Chapter 8 Measurement of Small Signals Floating Operation 8 8...
Страница 68: ...Appendix A Interface 1000 Specifications Floating Operation Appendix A 6...
Страница 70: ...Appendix B Interface 1000 Cell Connector Floating Operation Appendix B 2...
Страница 78: ...Appendix E CE Certificate Appendix E 2 Low Voltage Certificate of Conformance...
Страница 79: ...Appendix E CE Certificate Appendix E 3 RFI Certificate of Conformance...
Страница 80: ...Appendix E CE Certificate Appendix E 4...
Страница 84: ...Appendix F Heat in Interface 1000 MultEchem Systems Appendix F 4...