Appendix A: Interface 1000 Specifications--Floating Operation
Appendix A - 3
Potentiostatic Mode
Applied Voltage Range
Max
±12
volts
Accuracy
DC zero offset
Gain
Max
1
0.2
mV
% setting
Note 18
DC Bias
Typ
±8
volts
Scan DAC ranges
Typ
±6.4, ±1.6, ±0.4
volts
Drift
Max
<20
µV/
C
Note 19
Noise and Ripple
1 Hz to 1 kHz
1 Hz to 200 kHz
Typ
10
20
µV rms
Note 20
Galvanostatic Mode
Maximum Full-scale Current
±1000
mA
Minimum Full-scale Current
10
nA
Accuracy
Dominated by current to voltage
converter accuracy shown above
Sig Gen Voltage for Full-scale Current
±3.0
V
Environmental
Operating Temperature Range
0 to 45
C
Relative Humidity
Max
90 (non-condensing)
%
Storage and Shipping Temperature
–25 to 75
C
Maximum Shipping Acceleration
30
G
Содержание Interface 1000
Страница 8: ......
Страница 18: ......
Страница 36: ...Chapter 4 Cell Connections Membrane Cell Connections 4 4...
Страница 44: ...Chapter 5 Panel Indicators and Connectors Rear Panel 5 8...
Страница 62: ...Chapter 8 Measurement of Small Signals Floating Operation 8 8...
Страница 68: ...Appendix A Interface 1000 Specifications Floating Operation Appendix A 6...
Страница 70: ...Appendix B Interface 1000 Cell Connector Floating Operation Appendix B 2...
Страница 78: ...Appendix E CE Certificate Appendix E 2 Low Voltage Certificate of Conformance...
Страница 79: ...Appendix E CE Certificate Appendix E 3 RFI Certificate of Conformance...
Страница 80: ...Appendix E CE Certificate Appendix E 4...
Страница 84: ...Appendix F Heat in Interface 1000 MultEchem Systems Appendix F 4...