Appendix A: EIS Box Specifications
44
Current to Voltage Converter
Maximum Full-scale Range
±5
A
Note 13
Minimum Full-scale Range
±50
±500 (after 100× gain)
µA
nA
Note 13
Voltage across Rm
Typ
±100
mV at
full scale
Note 14
Output Voltage (at Monitor and ADC in)
Typ
±3.0
V
full scale
Input Offset Current
Max
25
pA
Note 15
Range Zero Offset
Max
0.05
% of
range
Note 15
Gain Tolerance, 5 A to 50 µA
ranges
Max
0.2
% of
reading
Note 15
Zero drift
Typ
0.02
% FS/
C
Note 16
Bandwidth (-3 dB)
Typ
5
MHz
Note 17
Current Measurement
Resolution
Typ
0.00333…
% FS/bit
Offset Range
Typ
±100
% of
range
Post-offset Gain
Typ
1×, 10×, 100×
Accuracy
Typ
Dominated by current-to-voltage error
(see above)
Environmental
Operating Temperature Range
0 to +45
°C
Relative Humidity
Max
90 (non-condensing)
%
Storage and Shipping Temperature
–
25 to +75
°C
Maximum Shipping Acceleration
30
g
General
DC Power Input Voltage
Range 100
–
240
V
Fuses
IEC 127
For 115 VAC, use 6.3 A 250 V
For 230 VAC, use 3.153 A 250 V
Power
Max
120
W
Leakage Current (floating, earthed
working electrode)
Typ
±10
nA
Note 22
Dimensions
41.5 W × 45.7 D × 13.3 H
cm
Mounting
Fits 19” (48.3 cm) standard rack mount
3 U height
Weight
Max
10
kg
Содержание EIS Box
Страница 2: ...2...
Страница 6: ...6...
Страница 14: ...Safety Considerations 14...
Страница 20: ...20...
Страница 28: ......
Страница 49: ...Appendix C CE Certificate 49 Appendix C CE Certificate Certificate of Conformance...
Страница 50: ...50...
Страница 52: ...52 Work Shield 49 Working electrode 33 34 35 39 49...