Gamry Instruments EIS Box Скачать руководство пользователя страница 33

Cell Connections  

33 

 

Table 6-2 

Cell Cable Connections for a Membrane Cell 

Color 

Type 

Name 

Normal Connection 

Blue 

Banana Plug 

Working Sense 

Connect to reference electrode #1 

Green 

Banana Plug 

Working Electrode 

Connect to counter electrode #1 

White 

Banana Plug 

Reference 

Connect to reference electrode #2 

Red 

Banana Plug 

Counter Electrode 

Connect to counter electrode #2 

 

 

Fuses in the Cell Cable 

The EIS Box can be damaged if currents much larger than 6 A flow into or out of the Counter electrode or 
Working electrode leads. Improper connection to a battery, fuel cell, or supercapacitor can cause this type of 
damaging current to flow. 

All standard EIS Box Counter/Working cell cables include fuses in the current-carrying leads. These fuses protect 
the instrument from the extremely large currents that can flow through an improperly connected 
electrochemical-energy generation or storage device (including batteries, fuel cells, and capacitors). For 
convenience, the term 

battery

” is

 used here to refer to all single-cell or stacked electrochemical devices that 

can distribute energy. 

Grounding errors on a battery can be particularly dangerous, for they can result in the battery being shorted 
through the instrument. 

 

Both the counter electrode lead and the working electrode lead have fuses, rated to carry 6.3 A. 

 

Access to the fuses requires removal of two screws on opposite sides of the hood covering the D-connector end 
of the Counter/Working cable. A drawing of the D-connector end of a cable with the hood opened is given in 
Figure 6-3. The small rectangular fuses snap into fuse-holders labeled 

Work

 and 

Cntr

. To remove the fuse, 

 

Warning:

 The fuses in the EIS Box Counter/Working cell cable do not protect against a 

safety hazard. They are needed to prevent damage to the instrument if it is improperly connected. 

 Reference electrode #1 and counter electrode #1 must be on one side of the membrane, 

and reference electrode #2 and counter electrode #2 must be on the other side. 

 Always turn off the EIS Box and disconnect both ends of the Counter/Working cable before 

checking or replacing the fuses in the cable. 

Содержание EIS Box

Страница 1: ...EIS Box Instrument Operator s Manual Copyright 2018 Gamry Instruments Inc Revision 1 0 November 1 2018 Gamry P N 988 00071...

Страница 2: ...2...

Страница 3: ...in installation or use of a system containing an EIS Box please try to call from a phone next to your computer where you can type and read the screen while talking to us We will be happy to provide a...

Страница 4: ...ny time without incurring any obligation to install same on systems previously purchased All system specifications are subject to change without notice There are no warranties which extend beyond the...

Страница 5: ...truments Inc assumes no responsibility for errors that might appear Copyrights EIS Box Instrument Operator s Manual copyright 2018 Gamry Instruments Inc all rights reserved Gamry Framework copyright 1...

Страница 6: ...6...

Страница 7: ...Cell Control 19 Communications Overview 19 Chapter 4 Installation 21 Initial Visual Inspection 21 Physical Location 21 Computer Requirements 22 Quick Start Guide for System Installation 22 Software I...

Страница 8: ...erential Electrometer 43 Voltage Measurement 43 Current to Voltage Converter 44 Current Measurement 44 Environmental 44 General 44 Channel Characteristics 45 Appendix B EIS Box Cell Connectors 47 Appe...

Страница 9: ...cuitry does not contain voltages higher than 15 V DC Generally input and output voltages in the EIS Box are limited to 9 V This voltage level is considered safe The EIS Box is normally provided with a...

Страница 10: ...y and other surge events such as lightning Warning Never connect the Protective Ground and System Ground if the electrochemical system you are testing is earth grounded Two earth ground connections in...

Страница 11: ...l Stresses Treat your EIS Box as potentially hazardous if any of the following is true of the unit It shows visible damage It does not operate properly It has been stored for a long period of time und...

Страница 12: ...se isopropyl alcohol Do not use a wet rag or allow fluid to enter the EIS Box enclosure Do not immerse the EIS Box in any type of cleaning fluid including water Do not use any abrasive cleaners Servic...

Страница 13: ...ectrical equipment for measurement control and laboratory use EMC Requirements describing acceptable limits for electrical transient susceptibility in laboratory test equipment It should continue to o...

Страница 14: ...Safety Considerations 14...

Страница 15: ...esigned to for battery testing via electrochemical impedance spectroscopy up to eight cells The EIS Box excels at electrochemical impedance testing of small to medium sized single cell energy storage...

Страница 16: ...t software Software and Applications The EIS Box supports Gamry Instruments electrochemical impedance spectroscopy software under the Framework system Typical applications for the EIS Box include Rese...

Страница 17: ...ts from other general purpose switching devices The EIS Box is integrated into all Gamry Instruments electrochemical impedance spectroscopy applications that run under the Gamry Instruments Framework...

Страница 18: ...ritical component of the system It controls the EIS Box by sending it simple commands through an ethernet or USB connection The most important function of these commands is to select which cell is act...

Страница 19: ...l An EIS Box contains eight channels A channel includes cell switching relays The term cell is used when only the cell switching portion of a channel is being discussed Inactive Cell The term inactive...

Страница 20: ...20...

Страница 21: ...o the rear of the instrument because some cable connections are made from the rear The EIS Box is generally operated in a flat position see Figure 4 1 Operation in other orientations is possible as lo...

Страница 22: ...ation so the information provided below is always subject to change Software Installation The EIS Box is compatible with the Windows Plug Play configuration system Like most Plug Play hardware it is b...

Страница 23: ...r of the EIS Box Fig 4 2 The power supply provided inside the EIS Box is rated for operation from 100 to 240 V AC at frequencies from 47 to 63 Hz It should therefore be useable worldwide The EIS Box i...

Страница 24: ...status of the other LED indicators is not important at this time Ethernet Cables The EIS Box connects to the computer using a shielded CAT5E Ethernet cable see figure below The Ethernet connection sho...

Страница 25: ...USB cable Cell Cable Installation The Cell Cables connect pairwise to two 15 pin D type connectors located on the front of the EIS Box see Figure 4 1 for each channel There are two cables per channel...

Страница 26: ...brate instruments connected to the system Framework Device Status Bar By default the Gamry Framework shows a Device Status Bar under its main menu see Figure 4 3 If you do not see the Device Status Ba...

Страница 27: ...dles most of the functions of the EIS Box The second is the Communications Firmware This program handles the communications between your EIS Box and the host computer The third is the PLD firmware whi...

Страница 28: ......

Страница 29: ...em A calibration utility is provided with the Gamry Framework The calibration for the EIS Box is divided into two sections Instrument DC Calibration and cable calibration Gain access to these calibrat...

Страница 30: ...been serviced You notice breaks or discontinuities in the data curves recorded with your system The system is being run in an environment that is very different from the previous operating environmen...

Страница 31: ...ery EIS Box is shipped with eight pairs of standard shielded cell cables The Gamry part number for the Current Cable is 985 00183 and the Sense Cable is 985 00184 They are both 3 m complex cables with...

Страница 32: ...power amplifier If your cell is a typical glass laboratory cell all of the electrodes are isolated from earth ground In this case you may be able to reduce noise in your data by connecting the EIS Box...

Страница 33: ...ence the term battery is used here to refer to all single cell or stacked electrochemical devices that can distribute energy Grounding errors on a battery can be particularly dangerous for they can re...

Страница 34: ...resistance of a good fuse is very small Figure 6 3 Drawing of Fuse holder in the D Connector Hood Four replacement fuses should accompany every Counter Working cell cable shipped by Gamry Instruments...

Страница 35: ...Utilities Set a Voltage on the Framework software s menu bar You see a Set a Voltage dialog box similar to this Figure 6 4 Setup Dialog Box for the Set a Voltage Script In the Voltage field enter a Vo...

Страница 36: ...current readings are near zero and a red CA Overload indication may be seen at the bottom of the runner window If this test indicates an open fuse use the procedures described above to check both fus...

Страница 37: ...ff There is no AC mains supply connected to the rear panel Power In connector One of the internal DC power supplies has failed Cell Connectors The EIS Box has two sets of eight cell cable connectors f...

Страница 38: ...el LED is between each pair of Cell connectors The Channel LEDs glow amber whenever the EIS Box is actively applying voltage or current to the electrochemical cell attached to that channel Avoid touch...

Страница 39: ...you can connect to the Chassis Ground jack Use a wire to connect the banana plug to earth ground A black banana plug to banana plug lead is provided with your EIS Box You may find it useful when makin...

Страница 40: ...ess point to the protective earth ground in the EIS Box It is connected to the third wire in the AC power cord and to the EIS Box s metal chassis This binding post can be connected to an additional ea...

Страница 41: ...on the left side of the EIS Box s rear panel It can be identified by the line cord connector built into it 3 Find the small slot on the right side of the AC input module near the line voltage setting...

Страница 42: ...Make sure you have not inadvertently changed the AC line voltage setting which is displayed in a small window on the AC input module If the voltage shown is not the appropriate setting for the line v...

Страница 43: ...Amplifier Compliance Voltage Min 6 5 V Note 1 Output Current Min 5 A Note 2 Unity Gain Bandwidth Typ 1050 250 43 4 4 0 5 kHz Note 3 Slew Rate Typ 10 4 4 0 7 0 07 0 007 V s Note 4 Differential Electrom...

Страница 44: ...5 MHz Note 17 Current Measurement Resolution Typ 0 00333 FS bit Offset Range Typ 100 of range Post offset Gain Typ 1 10 100 Accuracy Typ Dominated by current to voltage error see above Environmental O...

Страница 45: ...2 G resistor switched into the input and measuring the voltage difference with and without the input resistance 7 The differential impedance is measured between the Reference and Work Sense inputs Th...

Страница 46: ...The specified I E Converter BW is measured at 5 mA I E Range 18 The total error in a voltage setting is Error DC Zero Offset Gain Voltage Setting For a 1 V signal the theoretical error can be as high...

Страница 47: ...Ground The potentiostat s floating ground Can be used to shield the cell if very low currents need to be measured Also used as a shield for the counter electrode cable 5 12 Counter Connected to the c...

Страница 48: ...the cell end of the cell cable 7 15 Work Sense Shield The shield for the work sense electrode input Driven to the same potential as Pin 8 Left open at the end of the cell cable 8 Work Sense Connected...

Страница 49: ...Appendix C CE Certificate 49 Appendix C CE Certificate Certificate of Conformance...

Страница 50: ...50...

Страница 51: ...fan speed 12 firmware update 29 Floating Ground binding post 11 41 42 Framework software 17 28 31 fuses 35 36 37 38 43 Gamry Framework 17 Gamry Instrument Manager 28 31 green cell lead 34 ground 49 50...

Страница 52: ...52 Work Shield 49 Working electrode 33 34 35 39 49...

Отзывы: